Suppr超能文献

使用八个带有和不带有在线加湿功能的麻醉回路在体外测量无效通气。

Wasted ventilation measured in vitro with eight anesthetic circuits with and without inline humidification.

作者信息

Coté C J, Petkau A J, Ryan J F, Welch J P

出版信息

Anesthesiology. 1983 Nov;59(5):442-6. doi: 10.1097/00000542-198311000-00015.

Abstract

Compression of gases (Boyle's law) and circuit compliance are major determinants of anesthesia circuit function. The materials of which circuits are constructed and the use of heated humidifiers may result in clinically important variations in delivered minute ventilation (VE) secondary to variations in compression volume. We examined eight anesthetic circuits both with and without a heated humidifier in an in vitro setting. Compression volume was determined with a large calibrated syringe. Circuit efficiency was determined by measuring VE at multiple peak inflation pressures (PIP) while using a pediatric ventilator with fixed VE, respiratory rate, fresh gas flow, and I/E ratio. As expected, both compression volume and delivered VE highly correlated with the type of circuit and the pressure at which it was examined (P less than 0.001). Mapleson D circuits had the lowest compression volume and were the most efficient circuits (P less than 0.0001). Pediatric circle systems were intermediate and adult circle systems had the largest compression volume and were the least efficient. Humidifiers uniformly increased compression volume. The following conclusions were drawn: 1) the anesthetic circuit, its material, and the pressure at which it operates are important determinants of circuit function; 2) humidifiers increase compression volume; 3) Mapleson D circuits had the lowest compression volume and therefore were the most efficient; 4) highly compliant adult circuits may result in compression volume losses that exceed the tidal volume of a pediatric ventilator; 5) humidifiers with low volume and rigid tubing should have the least effect on minute ventilation; and 6) highly compliant adult circuits when used in the care of infants and small children must be used with caution.

摘要

气体压缩(波义耳定律)和回路顺应性是麻醉回路功能的主要决定因素。回路的构建材料以及加热湿化器的使用可能会因压缩容积的变化而导致输送分钟通气量(VE)出现临床上的重要差异。我们在体外环境中对八个有无加热湿化器的麻醉回路进行了检查。使用一个大的校准注射器测定压缩容积。通过在使用具有固定VE、呼吸频率、新鲜气体流量和吸呼比的小儿呼吸机时,在多个峰值充气压力(PIP)下测量VE来确定回路效率。正如预期的那样,压缩容积和输送的VE均与回路类型及其检查时的压力高度相关(P<0.001)。Mapleson D回路的压缩容积最低,是效率最高的回路(P<0.0001)。小儿环路系统居中,成人环路系统的压缩容积最大且效率最低。湿化器均会增加压缩容积。得出以下结论:1)麻醉回路、其材料以及其运行的压力是回路功能的重要决定因素;2)湿化器会增加压缩容积;3)Mapleson D回路的压缩容积最低,因此效率最高;4)顺应性高的成人回路可能导致压缩容积损失超过小儿呼吸机的潮气量;5)容积小且管道硬的湿化器对分钟通气的影响应最小;6)在照顾婴儿和小儿时使用顺应性高的成人回路时必须谨慎。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验