机械通气中气体压缩的校正

Correction for gas compression in mechanical ventilators.

作者信息

Forbat A F, Her C

出版信息

Anesth Analg. 1980 Jul;59(7):488-93.

PMID:6773435
Abstract

Errors in measurements arising from compression of gases in a mechanical ventilator may be corrected by calculation based on the compressible volume of the ventilator, or by physical separation of expired from compressed gases. Using a Frumin valve to effect this separation, the volume compressed in the Bear I ventilator was found to be 5.15 ml/mm Hg, and its compressible volume to be 3790 ml (range 3760 to 3806 ml). The ratio of uncorrected/corrected tidal volume (termed the correction factor) was 1.229 +/- 0.014. When VD/VT was corrected using this factor and compared with that obtained by calculating VD/VT using PECO2 measurements in expired air collected from the Frumin valve, the difference between the two was not significant. It is concluded that the use of the Frumin valve is a practical method for determining the true dead space-to-tidal volume ratios and that the application of the correction factor is also reliable.

摘要

机械通气中因气体压缩产生的测量误差,可通过基于通气机可压缩容积的计算来校正,或通过将呼出气体与压缩气体物理分离来校正。使用弗鲁明阀实现这种分离,发现熊I型通气机中压缩的容积为5.15毫升/毫米汞柱,其可压缩容积为3790毫升(范围为3760至3806毫升)。未校正/校正潮气量的比值(称为校正因子)为1.229±0.014。当使用该因子校正VD/VT并与通过使用从弗鲁明阀收集的呼出气体中的PECO2测量值计算VD/VT所获得的值进行比较时,两者之间的差异不显著。结论是,使用弗鲁明阀是确定真正的死腔与潮气量比值的实用方法,并且校正因子的应用也是可靠的。