Jones R N, Barry A L, Bigelow J, Gavan T L, Thornsberry C
J Clin Microbiol. 1982 Jul;16(1):153-63. doi: 10.1128/jcm.16.1.153-163.1982.
Four laboratories participated in a three-phase study to evaluate the MICUR antimicrobial broth microdilution system (Boehringer Mannheim Diagnostics, Inc., Houston, Tex.). The dried-antimicrobial agent MICUR system was compared with a reference broth microdilution method (National Committee for Clinical Laboratory Standards) by using 304 recently isolated clinical strains and two collections of stock or challenge organisms. Of 7,092 minimum inhibitory concentration (MIC) datum pairs derived from the clinical isolates, 96.6% were within an acceptable (+/- 1 log2 dilution) range. MICUR MICs agreed with the reference broth microdilution method MICs in 95.3% of 6,840 MIC pair determinations performed on stock or challenge cultures. The MICUR intralaboratory reproducibility within +/- 1 log2 dilution step for the clinical isolates was 98.4%. The MICUR intralaboratory and interlaboratory reproducibilities for 26 stock cultures were 98.4 and 95.1%, respectively. For 180 challenge cultures (4,199 MIC pairs) which were included in the MICUR testing to provide a wide variety of antimicrobial susceptibility and resistance patterns, the results for 92.5% were in close agreement with the reference broth microdilution results. No specific resistance mechanism went unrecognized by this new commercial system. The MICUR system gives comparable MIC results when evaluated against the reference broth microdilution method, and it would be acceptable for use in clinical microbiology laboratories.
四个实验室参与了一项分三个阶段的研究,以评估MICUR抗菌肉汤微量稀释系统(德国宝灵曼诊断公司,得克萨斯州休斯顿)。通过使用304株最近分离出的临床菌株以及两组标准菌株或质控菌株,将干燥抗菌剂MICUR系统与参考肉汤微量稀释法(美国国家临床实验室标准委员会)进行了比较。在从临床分离株获得的7092个最低抑菌浓度(MIC)数据对中,96.6%在可接受范围(±1个log2稀释度)内。在对标准菌株或质控培养物进行的6840次MIC对测定中,95.3%的MICUR MIC结果与参考肉汤微量稀释法的MIC结果一致。临床分离株在±1个log2稀释度步骤内的MICUR实验室内重复性为98.4%。26株标准菌株培养物的MICUR实验室内和实验室间重复性分别为98.4%和95.1%。在MICUR测试中纳入的180株质控培养物(4199个MIC对),用于提供广泛的抗菌敏感性和耐药模式,92.5%的结果与参考肉汤微量稀释结果高度一致。这种新的商业系统没有遗漏任何特定的耐药机制。与参考肉汤微量稀释法相比,MICUR系统给出了可比的MIC结果,可用于临床微生物实验室。