Brunton W A, Steel J M, Percy-Robb I W
Clin Chim Acta. 1977 Mar 15;75(3):359-64. doi: 10.1016/0009-8981(77)90353-9.
The Reflomat System for rapid estimation of plasma or blood glucose concentration has been evaluated. The System gave a linear response throughout its analytical range and the recovery of glucose added to glucose-free plasma was 97-105%. Addition of sodium fluoride to plasma produced a 7-15% reduction in the estimated glucose concentration. Plasma glucose concentration estimated with the Reflomat agreed closely with results of a glucose oxidase and a hexokinase based method, and blood glucose concentration measured with the Reflomat agreed well with results of a glucose oxidase method.