el-Attar A M
Department of Anesthesia, Alexandria University, Egypt.
Middle East J Anaesthesiol. 1996 Feb;13(4):395-403.
The lowest tidal volume delivered by most of the pediatric anesthesia ventilators could be large for neonates, and may generate unacceptably high peak airway pressures. The effective tidal volume can however be reduced by connection of a small rubber bag between the ventilator and the expiratory limb of the Ayre's T-piece circuit to divide the delivered gas mixture between the added bag and the neonate's lungs. The delivered tidal volume by the ventilator can then be manipulated to achieve the desired peak inspiratory pressure; 18-20 cm H2O in neonates with normal lung compliance. The modification was found to be simple, safe and convenient.
大多数小儿麻醉呼吸机输送的最小潮气量对新生儿来说可能过大,并且可能产生高得无法接受的气道峰压。然而,通过在呼吸机和Ayre T形管回路的呼气支路之间连接一个小橡胶袋,将输送的气体混合物在附加的袋子和新生儿肺部之间分配,可以减少有效潮气量。然后可以调节呼吸机输送的潮气量,以达到所需的吸气峰压;肺顺应性正常的新生儿为18 - 20 cmH₂O。发现这种改进简单、安全且方便。