Cage B R, Schreiber K, Barnes C, Portnoy J
Midwest Research Institute, Kansas City, Missouri, USA.
Ann Allergy Asthma Immunol. 1996 Nov;77(5):401-6. doi: 10.1016/S1081-1206(10)63339-X.
A comparative evaluation of four air samplers was performed using bioaerosol collection in the outdoor environment.
Test samplers used included a Rotorod, a Kramer-Collins suction trap, an all-glass impinger (AGI-30), and a high-volume cyclonic liquid impinger (SpinCon). All samples were analyzed microscopically for spores and pollen. The two collectors providing a liquid sample (AGI-30 and SpinCon) also were analyzed for specific allergen content by enzyme-linked immunoassay.
The SpinCon collected a larger number of spores than the other devices. The number of spores collected by this unit per volume of air sampled was comparable to the AGI-30. The Rotorod and Kramer-Collins collected a lower number of spores per unit of air but collected a larger number of pollen grains per volume sampled. Alternaria allergens Alt a I and GP70 were collected by both liquid impingers; however, the SpinCon collected more Alt a I and the AGI-30 collected more GP70.
The SpinCon is a device that is capable of efficiently sampling a high volume of air and concentrating it in a form that can be analyzed for the presence of spores and fungal allergens. It is less useful for collecting intact pollen grains. Pollen allergen quantitation has not yet been performed on the SpinCon effluent.
在室外环境中使用生物气溶胶收集对四种空气采样器进行了比较评估。
所使用的测试采样器包括转子棒采样器、克莱默 - 柯林斯吸气阱、全玻璃冲击式采样器(AGI - 30)和大容量旋风式液体冲击式采样器(SpinCon)。所有样本均通过显微镜分析孢子和花粉。对于提供液体样本的两种采样器(AGI - 30和SpinCon),还通过酶联免疫吸附测定法分析其特定过敏原含量。
SpinCon采集的孢子数量比其他设备多。该装置每采样体积空气中采集的孢子数量与AGI - 30相当。转子棒采样器和克莱默 - 柯林斯采样器每单位空气采集的孢子数量较少,但每采样体积采集的花粉粒数量较多。两种液体冲击式采样器均采集到链格孢属过敏原Alt a I和GP70;然而,SpinCon采集到的Alt a I更多,而AGI - 30采集到的GP70更多。
SpinCon是一种能够高效采集大量空气并将其浓缩成可分析孢子和真菌过敏原存在形式的设备。它在收集完整花粉粒方面不太有用。尚未对SpinCon流出物进行花粉过敏原定量分析。