Gheber L, Priel Z
Department of Chemistry, Ben-Gurion University of the Negev, Beer-Sheva, Israel.
Biophys J. 1997 Jan;72(1):449-62. doi: 10.1016/S0006-3495(97)78686-7.
Photoelectric signals were created and used to investigate the features of the signals as a function of the ciliary beat parameters. Moreover, correlation between the simulated and the measured signals permitted measurement of the cilium beat parameters. The simulations of the signals were based on generation of a series of time-frozen top-view frames of an active ciliary area and determination of the amount of light passing through an observation area in each of these frames. All the factors that might contribute to the shape of the signals, namely, partial ciliary transmittance of light, three-dimensional ciliary beat (composed of recovery, effective, and pause parts), phase distribution on the ciliary surface, and the large number of cilia that contribute to the photoelectric signal, were taken into account in generation of the signals. Changes in the ciliary parameters influenced the shape of the photoelectric signals, and the different phases of the beat could not be directly and unequivocally identified in the signals. The degree of temporal asymmetry of the beat and the portion of the cycle occupied by the pause significantly influenced the shapes of both the lower and the upper parts of the signal and the slopes of the signal. Increases in the angle of the arc swept by the cilium during the effective stroke smoothed the signals and increased the duration of the upper part of the signal. The angle of the arc projected by the cilium onto the cell surface during the recovery stroke had minor effects on the signal's shape. Characteristics of the metachronal wave also influenced the signal's shape markedly. Decreases in ciliary spacing smoothed the signals, whereas ciliary length had a minor influence on the simulated photoelectric signals. Comparison of the simulated and the measured signals showed that the beat parameters of the best-fitting simulated signals converged to values that agree well with the accepted range of beat parameters in mucociliary systems.
产生光电信号并用于研究这些信号作为纤毛摆动参数函数的特征。此外,模拟信号与测量信号之间的相关性允许测量纤毛摆动参数。信号的模拟基于生成一系列活跃纤毛区域的时间冻结顶视图帧,并确定这些帧中每个帧通过观察区域的光量。在生成信号时考虑了所有可能影响信号形状的因素,即光的部分纤毛透射率、三维纤毛摆动(由恢复、有效和暂停部分组成)、纤毛表面的相位分布以及对光电信号有贡献的大量纤毛。纤毛参数的变化影响了光电信号的形状,并且在信号中无法直接明确地识别出摆动的不同阶段。摆动的时间不对称程度和暂停占据的周期部分显著影响了信号上下部分的形状以及信号的斜率。有效冲程期间纤毛扫过的弧度角度增加使信号平滑,并增加了信号上部的持续时间。恢复冲程期间纤毛投射到细胞表面的弧度角度对信号形状的影响较小。同步波的特征也对信号形状有显著影响。纤毛间距减小使信号平滑,而纤毛长度对模拟光电信号的影响较小。模拟信号与测量信号的比较表明,最佳拟合模拟信号的摆动参数收敛到与粘液纤毛系统中公认的摆动参数范围相符的值。