Medical Radiation Science, National Physical Laboratory, Hampton Road, Teddington, Middlesex, TW11 0LW, United Kingdom.
Medical Physics and Biomedical Engineering, University College London, Gower Street, London, WC1E 6BT, United Kingdom.
Sci Rep. 2020 Jun 3;10(1):9089. doi: 10.1038/s41598-020-65819-y.
High dose-rate radiotherapy, known as FLASH, has been shown to increase the differential response between healthy and tumour tissue. Moreover, Very High Energy Electrons (VHEEs) provide more favourable dose distributions than conventional radiotherapy electron and photon beams. Plane-parallel ionisation chambers are the recommended secondary standard systems for clinical reference dosimetry of electrons, therefore chamber response to these high energy and high dose-per-pulse beams must be well understood. Graphite calorimetry, the UK primary standard, has been employed to measure the dose delivered from a 200 MeV pulsed electron beam. This was compared to the charge measurements of a plane-parallel ionisation chamber to determine the absolute collection efficiency and infer the ion recombination factor. The dose-per-pulse measured by the calorimeter ranged between 0.03 Gy/pulse and 5.26 Gy/pulse, corresponding to collection efficiencies between 97% and 4%, respectively. Multiple recombination models currently available have been compared with experimental results. This work is directly applicable to the development of standard dosimetry protocols for VHEE radiotherapy, FLASH radiotherapy and other high dose-rate modalities. However, the use of secondary standard ionisation chambers for the dosimetry of high dose-per-pulse VHEEs has been shown to require large corrections for charge collection inefficiency.
高剂量率放疗,即 FLASH,已被证明可以增加健康组织和肿瘤组织之间的差异反应。此外,甚高能电子(VHEE)比传统放疗电子和光子束提供更有利的剂量分布。平行板电离室是电子临床参考剂量学的推荐二级标准系统,因此必须很好地了解这些高能和高剂量脉冲束对腔室的响应。石墨量热计,英国的主要标准,已被用于测量 200 MeV 脉冲电子束的传递剂量。这与平行板电离室的电荷测量进行了比较,以确定绝对收集效率并推断离子复合因子。量热计测量的脉冲剂量在 0.03 Gy/脉冲和 5.26 Gy/脉冲之间,相应的收集效率分别为 97%和 4%。目前可用的多种复合模型已与实验结果进行了比较。这项工作直接适用于 VHEE 放疗、FLASH 放疗和其他高剂量率模式的标准剂量学协议的制定。然而,已经表明,对于高剂量脉冲 VHEE 的剂量测量,使用二级标准电离室会导致电荷收集效率不足的大校正。