Buser D, Nydegger T, Oxland T, Cochran D L, Schenk R K, Hirt H P, Snétivy D, Nolte L P
Department of Oral Surgery, School of Dental Medicine, University of Berne, Freiburgstrasse 7, CH-3010 Berne, Switzerland.
J Biomed Mater Res. 1999 May;45(2):75-83. doi: 10.1002/(sici)1097-4636(199905)45:2<75::aid-jbm1>3.0.co;2-p.
The purpose of the present study was to evaluate the interface shear strength of unloaded titanium implants with a sandblasted and acid-etched (SLA) surface in the maxilla of miniature pigs. The two best documented surfaces in implant dentistry, the machined and the titanium plasma-sprayed (TPS) surfaces served as controls. After 4, 8, and 12 weeks of healing, removal torque testing was performed to evaluate the interface shear strength of each implant type. The results revealed statistically significant differences between the machined and the two rough titanium surfaces (p <.00001). The machined surface demonstrated mean removal torque values (RTV) between 0.13 and 0.26 Nm, whereas the RTV of the two rough surfaces ranged between 1.14 and 1.56 Nm. At 4 weeks of healing, the SLA implants yielded a higher mean RTV than the TPS implants (1.39 vs. 1. 14 Nm) without reaching statistical significance. At 8 and 12 weeks of healing, the two rough surfaces showed similar mean RTVs. The implant position also had a significant influence on removal torques for each implant type primarily owing to differences in density in the periimplant bone structure. It can be concluded that the interface shear strength of titanium implants is significantly influenced by their surface characteristics, since the machined titanium surface demonstrated significantly lower RTV in the maxilla of miniature pigs compared with the TPS and SLA surfaces.
本研究的目的是评估在小型猪上颌骨中未加载的具有喷砂和酸蚀(SLA)表面的钛种植体的界面剪切强度。种植体牙科学中记录最充分的两种表面,即机械加工表面和钛等离子喷涂(TPS)表面作为对照。在愈合4、8和12周后,进行去除扭矩测试以评估每种种植体类型的界面剪切强度。结果显示,机械加工表面与两个粗糙钛表面之间存在统计学上的显著差异(p <.00001)。机械加工表面的平均去除扭矩值(RTV)在0.13至0.26 Nm之间,而两个粗糙表面的RTV在1.14至1.56 Nm之间。在愈合4周时,SLA种植体的平均RTV高于TPS种植体(1.39对1.14 Nm),但未达到统计学显著性。在愈合8周和12周时,两个粗糙表面的平均RTV相似。种植体位置对每种种植体类型的去除扭矩也有显著影响,这主要是由于种植体周围骨结构密度的差异。可以得出结论,钛种植体的界面剪切强度受其表面特性的显著影响,因为与TPS和SLA表面相比,机械加工的钛表面在小型猪上颌骨中的RTV显著更低。