Phys Med Biol. 2005 Nov 7;50(21):N309-16. doi: 10.1088/0031-9155/50/21/N03. Epub 2005 Oct 12.

Abstract

Although not as widespread as Ir-192, Co-60 is also available on afterloading equipment devoted to high dose rate brachytherapy, mainly addressed to the treatment of gynaecological lesions. The purpose of this study is to obtain the dosimetric parameters of the Co-60 source used by the BEBIG MultiSource remote afterloader (BEBIG GmbH, Germany) for which there are no dosimetric data available in the literature. The Monte Carlo code GEANT4 has been used to obtain the TG43 parameters and the 2D dose rate table in Cartesian coordinates of the BEBIG Co-60 HDR source. The dose rate constant, radial dose function and anisotropy function have been calculated and are presented in a tabular form as well as a detailed 2D dose rate table in Cartesian coordinates. These dosimetric datasets can be used as input data and to validate the treatment planning system calculations.

摘要

虽然不像铱-192那样广泛使用,但钴-60也可用于专门用于高剂量率近距离放射治疗的后装设备,主要用于治疗妇科病变。本研究的目的是获取德国BEBIG GmbH公司的BEBIG多源遥控后装设备所使用的钴-60源的剂量学参数,而文献中尚无该剂量学数据。已使用蒙特卡罗代码GEANT4获取BEBIG钴-60高剂量率源的TG43参数和笛卡尔坐标下的二维剂量率表。已计算剂量率常数、径向剂量函数和各向异性函数,并以表格形式呈现,同时还给出了笛卡尔坐标下详细的二维剂量率表。这些剂量学数据集可用作输入数据,并用于验证治疗计划系统的计算。