在不同光照条件下测量对比敏感度:三种测试方法的比较。

Measuring contrast sensitivity under different lighting conditions: comparison of three tests.

作者信息

Bühren Jens, Terzi Evdoxia, Bach Michael, Wesemann Wolfgang, Kohnen Thomas

机构信息

Department of Ophthalmology, Johann Wolfgang Goethe-University, Frankfurt am Main, Germany.

出版信息

Optom Vis Sci. 2006 May;83(5):290-8. doi: 10.1097/01.opx.0000216100.93302.2d.

Abstract

PURPOSE

The purpose of this study was to evaluate three psychophysical tests for the measurement of contrast sensitivity (CS) and disability glare (DG) at different luminance levels.

METHODS

In 60 eyes of 60 individuals (group 1: 20 healthy eyes of young individuals; group 2: 20 healthy eyes of elderly subjects; group 3: 20 eyes with nuclear cataract), CS with best correction was measured twice with the Frankfurt-Freiburg Contrast and Acuity Test System (FF-CATS) and the Functional Acuity Contrast Test (FACT, 1.5 cycles per degree [cpd]) at 167 cd/m2 and 0.167 cd/m2, and with the Pelli-Robson Chart (PRC) at 100 cd/m2 with and without glare. Repeatability of test and retest, and discriminative ability between the different subgroups, were assessed for CS values.

RESULTS

Maximum CS values varied across tests. In all groups, highest CS values were obtained with the photopic FF-CATS. For FACT scores at 1.5 cpd, there was a ceiling effect for young subjects. CS scores obtained with the PRC were the lowest. The PRC had the best test-retest repeatability of all tests. Under mesopic conditions with glare, reliability was generally lower; the FF-CATS had the highest repeatability of the mesopic tests. The FF-CATS discriminated best between the different groups for all conditions.

CONCLUSIONS

There are large discrepancies in the test results between CS testing methods, especially under different lighting conditions. Results from different CS tests are not interchangeable.

摘要

目的

本研究旨在评估三种心理物理学测试,以测量不同亮度水平下的对比敏感度(CS)和失能眩光(DG)。

方法

对60名个体的60只眼睛进行研究(组1:20名年轻个体的健康眼睛;组2:20名老年受试者的健康眼睛;组3:20只核性白内障眼睛)。使用法兰克福 - 弗莱堡对比与视力测试系统(FF - CATS)和功能性视力对比测试(FACT,每度1.5周/度[cpd]),在167 cd/m²和0.167 cd/m²的亮度下,对最佳矫正视力下的CS进行两次测量,并使用佩利 - 罗布森视力表(PRC)在100 cd/m²有眩光和无眩光的情况下进行测量。评估了CS值的测试 - 重测重复性以及不同亚组之间的鉴别能力。

结果

不同测试的最大CS值有所不同。在所有组中,明视觉FF - CATS获得的CS值最高。对于1.5 cpd的FACT评分,年轻受试者存在天花板效应。PRC获得的CS评分最低。PRC在所有测试中具有最佳的测试 - 重测重复性。在有眩光的中间视觉条件下,可靠性通常较低;FF - CATS在中间视觉测试中具有最高的重复性。在所有条件下,FF - CATS对不同组的鉴别能力最佳。

结论

CS测试方法之间的测试结果存在很大差异,尤其是在不同的光照条件下。不同CS测试的结果不可互换。