Kim Hanna Y, Budenz Donald L, Lee Pak S, Feuer William J, Barton Keith
Department of Ophthalmology, Stanford University, Stanford, CA, USA.
Am J Ophthalmol. 2008 Feb;145(2):228-232. doi: 10.1016/j.ajo.2007.09.030. Epub 2007 Dec 11.
To determine if there is a systematic difference in central corneal thickness (CCT) measured using anterior segment optical coherence tomography (AS-OCT) as compared with ultrasound pachymetry.
Prospective observational study.
Consecutive subjects with clinically normal corneas underwent CCT measurement by both ultrasound and AS-OCT while participating in a population-based study in Ghana, West Africa. One eye of each subject was randomly selected for analysis. Two measurements were taken and averaged. Agreement and interobserver variability were also analyzed.
One hundred and fifty-five subjects of African ethnicity and average age 57 years (standard deviation [SD] 12; range, 40 to 98 years) were included. Measurements by AS-OCT and US were taken a mean of 15 days (maximum, six weeks) apart. The mean (SD) [range] US CCT was 525.3 microm (33.5) [422, 653] and 499.0 microm (32.0) [428, 613] with AS-OCT. Measurements by the two modalities were strongly correlated (r(2) = 0.82; P < .001), and a significant difference was observed between mean US and AS-OCT CCT (SD) [range] of 26.3 microm (14.2) [-63, 12] (P < .001). The width of the limits of agreement was 28 microm, about 6% of the average pachymetry reading. In 50 eyes randomly remeasured with OCT by a second observer, the intraclass correlation coefficient was 0.91. There was a small but significant systematic difference between observers (mean 6.9 microm, SD 10.9 microm), or 1.4% (P < .001), increasing the difference noted above.
There is a reproducible systematic difference between CCT measurements taken with ultrasound and OCT. It is important to note in clinical practice, that measurements acquired by these two modalities are not directly interchangeable.
确定使用眼前节光学相干断层扫描(AS-OCT)测量的中央角膜厚度(CCT)与超声角膜测厚法相比是否存在系统差异。
前瞻性观察研究。
在西非加纳进行的一项基于人群的研究中,连续入选临床上角膜正常的受试者,同时采用超声和AS-OCT测量CCT。为每个受试者随机选择一只眼睛进行分析。进行两次测量并取平均值。还分析了一致性和观察者间的变异性。
纳入155名非洲裔受试者,平均年龄57岁(标准差[SD]12;范围40至98岁)。AS-OCT和超声测量之间平均间隔15天(最长六周)。超声测量的CCT平均值(SD)[范围]为525.3微米(33.5)[422, 653],AS-OCT测量的为499.0微米(32.0)[428, 613]。两种测量方法高度相关(r(2)= 0.82;P<0.001),超声和AS-OCT测量的CCT平均值(SD)[范围]存在显著差异,为26.3微米(14.2)[-63, 12](P<0.001)。一致性界限宽度为28微米,约为平均角膜测厚读数的6%。在由第二位观察者用OCT对50只眼睛进行随机复测时,组内相关系数为0.91。观察者之间存在小但显著的系统差异(平均值6.9微米,SD 10.9微米),即1.4%(P<0.001),这增加了上述差异。
超声和OCT测量的CCT之间存在可重复的系统差异。在临床实践中需要注意的是,这两种方法获得的测量结果不能直接互换。