Infrared tympanic thermometry in a hot environment.

作者信息

Coso J D, Aguado-Jimenez R, Mora-Rodriguez R

机构信息

Facultad de Ciencias del Deporte, Universidad de Castilla-La Mancha, Toledo, Spain.

出版信息

Int J Sports Med. 2008 Sep;29(9):713-8. doi: 10.1055/s-2007-989417. Epub 2008 Jan 22.

Abstract

The purpose of this study was to compare tympanic (measured by infrared thermometry; Tty- (IRED)) with rectal and esophageal temperatures (T(REC) and T(ES)) during exercise in the heat. During Experiment 1, nine subjects pedaled for 55 min in a hot-dry environment (37 degrees C; 27% humidity) in still-air (<0.2 m/s) and for 10 additional min using water ingestion, wind and ice to cool them down. During Experiment 2, subjects pedaled for 90 min in a similar environment but with airflow at 2.5 m/s. Pearson correlation coefficients (r) and Bland-Altman plots were calculated. In Experiment 1, Tty-(IRED) and T (REC) were highly correlated (r=0.83; p<0.001) with close agreement (-0.08+/-0.4 degrees C). Overall Tty-(IRED) was significantly correlated with T(ES) (r=0.91; p<0.001). Cold water ingestion did not affect Tty-(IRED) or T(REC) but lowered T(ES). Wind and ice application lowered Tty- (IRED) below T(REC) (p<0.05). During Experiment 2, Tty-(IRED) was lower than T(REC) (p<0.05) and the difference increased throughout exercise as hyperthermia developed resulting in low agreement (-1.01+/-1.1 degrees C). In conclusion, Tty- (IRED) dangerously underestimates T(REC) when exercising in a hot environment with airflow or during a cooling treatment. However, it could correctly detect hyperthermia during exercise in a hot still-air environment.

摘要

本研究的目的是比较热环境下运动期间鼓膜温度(通过红外测温法测量;Tty-(IRED))与直肠温度和食管温度(T(REC)和T(ES))。在实验1中,9名受试者在炎热干燥环境(37摄氏度;27%湿度)的静止空气中(风速<0.2米/秒)蹬车55分钟,随后使用饮水、吹风和冰敷进行额外10分钟的降温。在实验2中,受试者在类似环境中蹬车90分钟,但风速为2.5米/秒。计算了Pearson相关系数(r)和Bland-Altman图。在实验1中,Tty-(IRED)与T(REC)高度相关(r = 0.83;p<0.001),一致性良好(-0.08±0.4摄氏度)。总体而言,Tty-(IRED)与T(ES)显著相关(r = 0.91;p<0.001)。饮用冷水未影响Tty-(IRED)或T(REC),但降低了T(ES)。吹风和冰敷使Tty-(IRED)低于T(REC)(p<0.05)。在实验2中,Tty-(IRED)低于T(REC)(p<0.05),且随着运动中体温过高的发展,差异在整个运动过程中增大,导致一致性较差(-1.01±1.1摄氏度)。总之,在有气流的炎热环境中运动或进行降温治疗时,Tty-(IRED)会危险地低估T(REC)。然而,它可以在炎热静止空气环境中的运动期间正确检测体温过高。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验