Hammer Daniel, Ferguson R Daniel, Iftimia Nicusor, Ustun Teoman, Wollstein Gadi, Ishikawa Hiroshi, Gabriele Michelle, Dilworth William, Kagemann Larry, Schuman Joel
Opt Express. 2005 Oct 3;13(20):7937-47. doi: 10.1364/opex.13.007937.
An upgraded optical coherence tomography system with integrated retinal tracker (TOCT) was developed. The upgraded system uses improved components to extend the tracking bandwidth, fully integrates the tracking hardware into the optical head of the clinical OCT system, and operates from a single software platform. The system was able to achieve transverse scan registration with sub-pixel accuracy (~10 microm). We demonstrate several advanced scan sequences with the TOCT, including composite scans averaged (co-added) from multiple B-scans taken consecutively and several hours apart, en face images collected by summing the A-scans of circular, line, and raster scans, and three-dimensional (3D) retinal maps of the fovea and optic disc. The new system achieves highly accurate OCT scan registration yielding composite images with significantly improved spatial resolution, increased signal-to-noise ratio, and reduced speckle while maintaining well-defined boundaries and sharp fine structure compared to single scans. Precise re-registration of multiple scans over separate imaging sessions demonstrates TOCT utility for longitudinal studies. En face images and 3D data cubes generated from these data reveal high fidelity image registration with tracking, despite scan durations of more than one minute.
开发了一种带有集成视网膜追踪器的升级版光学相干断层扫描系统(TOCT)。升级版系统采用了改进的组件来扩展追踪带宽,将追踪硬件完全集成到临床光学相干断层扫描(OCT)系统的光学头中,并在单一软件平台上运行。该系统能够实现亚像素精度(约10微米)的横向扫描配准。我们展示了使用TOCT的几种先进扫描序列,包括从连续采集且间隔数小时的多个B扫描平均(叠加)得到的复合扫描、通过对圆形、直线和光栅扫描的A扫描求和收集的正面图像,以及中央凹和视盘的三维(3D)视网膜图谱。与单次扫描相比,新系统实现了高精度的OCT扫描配准,生成的复合图像具有显著提高的空间分辨率、增加的信噪比和减少的散斑,同时保持清晰的边界和锐利的精细结构。在单独的成像会话中对多次扫描进行精确的重新配准证明了TOCT在纵向研究中的实用性。尽管扫描持续时间超过一分钟,但从这些数据生成的正面图像和3D数据立方体显示了带有追踪的高保真图像配准。