Cornea. 2012 Sep;31(9):983-8. doi: 10.1097/ICO.0b013e31823f8e2f.
To compare central corneal thickness (CCT) and anterior chamber depth (ACD) measured using 3 different techniques.
CCT and ACD were measured in one eye of 27 healthy subjects (age, mean ± SD: 34 ± 7 years), using the LenStar LS900, Pentacam, and Visante AS-OCT. The agreement between the measurement techniques was assessed using the coefficient of agreement (CoA) and the 95% limits of agreement (LoA).
Mean ± SD values for CCT using the LenStar, Pentacam, and Visante were 550.7 ± 37.4 μm, 542.7 ± 37.9 μm, and 556.7 ± 44.4 μm, respectively (P = 0.005). Plots of differences against means displayed relatively good agreement between the Pentacam and LenStar (CoA, 19.97 μm; LoA, 15.53 μm to -24.40 μm), poorer agreement between the LenStar and Visante (CoA, 31.26 μm; LoA, 40.78 μm to -21.74 μm), and agreement was poor between the Pentacam and Visante (CoA, 37.36 μm; LoA, 25.61 μm to -49.11 μm). Mean ± SD values for ACD using the LenStar, Pentacam, and Visante were 2.93 ± 0.30 mm, 2.96 ± 0.32 mm, and 3.03 ± 0.29 mm, respectively (P < 0.001). Relatively good agreement for ACD measures was observed between the Pentacam and LenStar (CoA, 0.08 mm; LoA, 0.12 to -0.04 mm), poor agreement was obtained between the Visante and LenStar (CoA, 0.13 mm; LoA, 0.21 to -0.06 mm), and agreement was also poor between the Pentacam and Visante (CoA, 0.14 mm; LoA, 0.11 to -0.17 mm).
Our data show that the LenStar, Pentacam, and Visante provide measurements that are in agreement with published values for CCT and ACD in human subjects. Although reasonable agreement for CCT and ACD was found between the Pentacam and LenStar, agreement was poorer between the Visante and LenStar and between the Visante and Pentacam. Thus, CCT and ACD measures from these instruments should not be used interchangeably.
比较三种不同技术测量的中央角膜厚度(CCT)和前房深度(ACD)。
对 27 名健康受试者(年龄,平均值±标准差:34±7 岁)的一只眼睛进行 CCT 和 ACD 测量,使用 LenStar LS900、Pentacam 和 Visante AS-OCT。使用一致性系数(CoA)和 95%一致性界限(LoA)评估测量技术之间的一致性。
LenStar、Pentacam 和 Visante 测量的 CCT 平均值±标准差分别为 550.7±37.4μm、542.7±37.9μm 和 556.7±44.4μm(P=0.005)。差值与平均值的图显示 Pentacam 和 LenStar 之间具有相对较好的一致性(CoA,19.97μm;LoA,15.53μm 至-24.40μm),LenStar 和 Visante 之间的一致性较差(CoA,31.26μm;LoA,40.78μm 至-21.74μm),Pentacam 和 Visante 之间的一致性较差(CoA,37.36μm;LoA,25.61μm 至-49.11μm)。LenStar、Pentacam 和 Visante 测量的 ACD 平均值±标准差分别为 2.93±0.30mm、2.96±0.32mm 和 3.03±0.29mm(P<0.001)。Pentacam 和 LenStar 之间观察到 ACD 测量值具有相对较好的一致性(CoA,0.08mm;LoA,0.12 至-0.04mm),Visante 和 LenStar 之间的一致性较差(CoA,0.13mm;LoA,0.21 至-0.06mm),Pentacam 和 Visante 之间的一致性也较差(CoA,0.14mm;LoA,0.11 至-0.17mm)。
我们的数据表明,LenStar、Pentacam 和 Visante 提供的 CCT 和 ACD 测量值与人体受试者的已发表值一致。尽管 Pentacam 和 LenStar 之间的 CCT 和 ACD 有合理的一致性,但 Visante 和 LenStar 之间以及 Visante 和 Pentacam 之间的一致性较差。因此,这些仪器的 CCT 和 ACD 测量值不应互换使用。