Reproducibility of retinal nerve fiber layer thickness measurements using spectral domain optical coherence tomography.

机构信息

Glaucoma Service, Massachusetts Eye and Ear Infirmary, Department of Ophthalmology, Harvard Medical School, Boston, MA 02114, USA.

出版信息

J Glaucoma. 2011 Oct;20(8):470-6. doi: 10.1097/IJG.0b013e3181f3eb64.

Abstract

PURPOSE

To evaluate the reproducibility of the peripapillary retinal nerve fiber layer (RNFL) thickness measurements obtained by Spectralis spectral domain optical coherence tomography (OCT) (Heidelberg Engineering, Heidelberg, Germany) in normal and glaucoma participants.

METHODS

Participants were recruited from a university-based clinic. Peripapillary RNFL thickness measurements were repeated 3 times during the same visit using the follow-up function. One eye of each participant was randomly selected for statistical analysis. Reproducibility was evaluated using within-subject standard deviation (Sw), coefficient of variation (CV), and intraclass correlation coefficient (ICC). Spearman rank correlation coefficient analyses were used to assess the correlation of the standard deviation of the 3 measurements for each participant with the RNFL thickness value.

RESULTS

Forty-five normal participants and 33 glaucoma patients were included in the study. The CVs ranged from 1.45% [overall global (G)] to 2.59% [temporal quadrant (T)] in normal eyes and from 1.74% (G) to 3.22% (T) in the glaucomatous eyes. ICCs ranged from 0.977 (T) to 0.990 (G and inferior-nasal sector) in normal eyes and from 0.983 (T) to 0.997 (inferior quadrant) in glaucomatous eyes. Sw were from 1.34 μm (G) to 2.39 μm (superior-temporal and inferior-temporal sectors) in normal eyes and from 1.14 μm (G) to 2.25 μm (superior-nasal sector) in the glaucomatous eyes. There were no significant correlations between RNFL thickness values and the measurement variability for each participant.

CONCLUSIONS

Spectralis OCT shows excellent reproducibility for measuring the peripapillary RNFL thickness values in both healthy and glaucoma participants.

摘要

目的

评估海德堡光谱光学相干断层扫描仪(Spectralis,德国海德堡工程公司)检测正常人和青光眼患者视盘周围视网膜神经纤维层(RNFL)厚度的重复性。

方法

参与者从大学诊所招募。使用随访功能在同一次就诊时重复测量 3 次视盘周围 RNFL 厚度。每位参与者的一只眼被随机选择进行统计分析。使用组内标准差(Sw)、变异系数(CV)和组内相关系数(ICC)评估重复性。Spearman 秩相关系数分析用于评估每个参与者的 3 次测量标准差与 RNFL 厚度值的相关性。

结果

本研究纳入 45 名正常人和 33 名青光眼患者。正常眼的 CV 范围为 1.45%(整体全局[G])至 2.59%(颞区[T]),青光眼眼的 CV 范围为 1.74%(G)至 3.22%(T)。正常眼的 ICC 范围为 0.977(T)至 0.990(G 和下鼻象限),青光眼眼的 ICC 范围为 0.983(T)至 0.997(下象限)。正常眼的 Sw 范围为 1.34μm(G)至 2.39μm(上颞和下颞区),青光眼眼的 Sw 范围为 1.14μm(G)至 2.25μm(上鼻区)。RNFL 厚度值与每个参与者的测量变异性之间无显著相关性。

结论

Spectralis OCT 对测量健康人和青光眼患者视盘周围 RNFL 厚度具有极好的重复性。

相似文献

引用本文的文献

2
Retinal manifestations of traumatic brain injury.创伤性脑损伤的视网膜表现
Sci Rep. 2025 Apr 29;15(1):14992. doi: 10.1038/s41598-025-94091-1.

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验