用自发或实验性青光眼的小鼠对 TonoLab 眼压计进行校准。

Calibration of the TonoLab tonometer in mice with spontaneous or experimental glaucoma.

机构信息

Glaucoma Service, Wilmer Ophthalmological Institute, The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.

出版信息

Invest Ophthalmol Vis Sci. 2011 Feb 22;52(2):858-64. doi: 10.1167/iovs.10-5556.

Abstract

PURPOSE

To measure the accuracy of TonoLab (TioLat, Helsinki, Finland) tonometry in mice with spontaneous or induced experimental glaucoma.

METHODS

Chronic intraocular pressure (IOP) elevation was induced in one eye of 32 mice by injection of polystyrene beads and viscoelastic material. Three to 6 weeks later, the eyes were cannulated and manometrically set to 10, 20, 30, 40, or 50 mm Hg. The mice were 8-week and 8-month-old C57BL/6, 8-week-old DBA/2J, and 8-week-old CD1. The TonoLab calibration was also tested on five aged DBA/2J mice with spontaneous glaucoma. The relation of the TonoLab reading to manometric IOP was evaluated in multivariate linear regression models with axial length, IOP history, and mouse strain as independent variables.

RESULTS

The slope of the relationship between TonoLab and manometric IOP in all the mice was 0.998, with an intercept of 2.3 mm Hg (adjusted R in univariate regression = 0.86). Neither the mice with bead-induced glaucoma nor those with spontaneous glaucoma (older DBA/2J mice) differed significantly from the control animals in having an excellent correlation between TonoLab and manometer IOP. Longer and wider mouse eyes had slightly higher tonometrically measured IOP, whether glaucomatous or control (multivariate regression, adjusted R(2) = 0.90, P < 0.0001). There was no difference in tonometric accuracy among the three mouse strains: CD1, C57BL/6, and DBA/2J, nor between 8-week and 8-month-old C57BL/6 mice (multivariate regression, P = 0.32).

CONCLUSIONS

The TonoLab accurately reflects IOP in both normal mice and in eyes of mice with experimental or spontaneous glaucoma, with no detectable effect of age.

摘要

目的

测量 TonoLab(TioLat,芬兰赫尔辛基)眼压计在自发性或实验性青光眼小鼠中的准确性。

方法

通过注射聚苯乙烯珠和粘性材料在 32 只小鼠的一只眼内诱导慢性眼内压(IOP)升高。3 至 6 周后,对眼睛进行套管插入并进行 10、20、30、40 或 50mmHg 的压力测量。这些小鼠为 8 周和 8 月龄的 C57BL/6、8 周龄的 DBA/2J 和 8 周龄的 CD1。还在 5 只患有自发性青光眼的老年 DBA/2J 小鼠上测试了 TonoLab 的校准。使用多元线性回归模型,以眼轴长度、IOP 史和小鼠品系为自变量,评估 TonoLab 读数与眼压计 IOP 的关系。

结果

所有小鼠的 TonoLab 与眼压计 IOP 之间的关系斜率为 0.998,截距为 2.3mmHg(单变量回归中的调整 R2=0.86)。无论是珠粒诱导性青光眼的小鼠还是自发性青光眼(老年 DBA/2J 小鼠),它们的 TonoLab 与眼压计 IOP 之间的相关性都非常好,与对照组动物没有显著差异。较长和较宽的小鼠眼睛的眼压计测量值略高,无论是否患有青光眼(多元回归,调整 R2=0.90,P<0.0001)。三种小鼠品系(CD1、C57BL/6 和 DBA/2J)之间的眼压计准确性没有差异,8 周和 8 月龄的 C57BL/6 小鼠之间也没有差异(多元回归,P=0.32)。

结论

TonoLab 准确反映了正常小鼠和实验性或自发性青光眼小鼠眼睛中的 IOP,并且年龄没有可检测的影响。

相似文献

1
Calibration of the TonoLab tonometer in mice with spontaneous or experimental glaucoma.
Invest Ophthalmol Vis Sci. 2011 Feb 22;52(2):858-64. doi: 10.1167/iovs.10-5556.
3
Effect of Antiglaucoma Medicine on Intraocular Pressure in DBA/2J Mice.
Ophthalmic Res. 2016;55(4):205-11. doi: 10.1159/000444057. Epub 2016 Mar 5.
5
Noninvasive measurement of rodent intraocular pressure with a rebound tonometer.
Invest Ophthalmol Vis Sci. 2005 Dec;46(12):4617-21. doi: 10.1167/iovs.05-0781.
6
Reliability and sensitivity of the TonoLab rebound tonometer in awake Brown Norway rats.
Invest Ophthalmol Vis Sci. 2009 Jun;50(6):2802-8. doi: 10.1167/iovs.08-2465. Epub 2009 Mar 25.
7
Differential susceptibility to experimental glaucoma among 3 mouse strains using bead and viscoelastic injection.
Exp Eye Res. 2010 Sep;91(3):415-24. doi: 10.1016/j.exer.2010.06.018. Epub 2010 Jun 26.
9
Investigating the effect of ciliary body photodynamic therapy in a glaucoma mouse model.
Invest Ophthalmol Vis Sci. 2006 Jun;47(6):2498-507. doi: 10.1167/iovs.05-0959.

引用本文的文献

1
A Comprehensive Protocol for Microbead-Induced Ocular Hypertension in Mice.
Methods Mol Biol. 2025;2858:243-264. doi: 10.1007/978-1-0716-4140-8_20.
2
Differential protection by nicotinamide in a mouse model of glaucoma DBA/2J revealed by second-harmonic generation microscopy.
PLoS One. 2024 Sep 10;19(9):e0309400. doi: 10.1371/journal.pone.0309400. eCollection 2024.
4
Effects of chronic mild hyperoxia on retinal and choroidal blood flow and retinal function in the DBA/2J mouse model of glaucoma.
PLoS One. 2022 Mar 25;17(3):e0266192. doi: 10.1371/journal.pone.0266192. eCollection 2022.
5
Safety, efficacy and delivery of multiple nucleoside analogs via drug encapsulated carbon (DECON) based sustained drug release platform.
Eur J Pharm Biopharm. 2022 Apr;173:150-159. doi: 10.1016/j.ejpb.2022.03.001. Epub 2022 Mar 21.
9
Retinoprotective Effects of PACAP Eye Drops in Microbead-Induced Glaucoma Model in Rats.
Int J Mol Sci. 2021 Aug 17;22(16):8825. doi: 10.3390/ijms22168825.
10
Intraocular Pressure Elevation Compromises Retinal Ganglion Cell Light Adaptation.
Invest Ophthalmol Vis Sci. 2020 Oct 1;61(12):15. doi: 10.1167/iovs.61.12.15.

本文引用的文献

1
Differential susceptibility to experimental glaucoma among 3 mouse strains using bead and viscoelastic injection.
Exp Eye Res. 2010 Sep;91(3):415-24. doi: 10.1016/j.exer.2010.06.018. Epub 2010 Jun 26.
2
The microbead occlusion model: a paradigm for induced ocular hypertension in rats and mice.
Invest Ophthalmol Vis Sci. 2010 Jan;51(1):207-16. doi: 10.1167/iovs.09-3947. Epub 2009 Oct 22.
3
Mouse models of retinal ganglion cell death and glaucoma.
Exp Eye Res. 2009 Apr;88(4):816-24. doi: 10.1016/j.exer.2008.12.002. Epub 2008 Dec 7.
4
Assessing intraocular pressure by rebound tonometer in rats with an air-filled anterior chamber.
Jpn J Ophthalmol. 2008 Nov-Dec;52(6):500-503. doi: 10.1007/s10384-008-0591-1. Epub 2008 Dec 17.
8
Dependency of intraocular pressure elevation and glaucomatous changes in DBA/2J and DBA/2J-Rj mice.
Invest Ophthalmol Vis Sci. 2008 Feb;49(2):613-21. doi: 10.1167/iovs.07-0745.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验