Suppr超能文献

新型非接触式光学低相干反射仪在儿童眼部生物测量中的可重复性

Reproducibility of ocular biometry with a new noncontact optical low-coherence reflectometer in children.

作者信息

Şahin Afsun, Gürsoy Hüseyin, Başmak Hikmet, Yildirim Nilgün, Usalp Zuhat, Çolak Ertuğrul

机构信息

Department of Ophthalmology, Eskişehir Osmangazi University Medical Faculty, Eskişehir, Turkey.

出版信息

Eur J Ophthalmol. 2011 Mar-Apr;21(2):194-8. doi: 10.5301/ejo.2010.1298.

Abstract

PURPOSE

To evaluate the reproducibility of Lenstar in the pediatric population.

METHODS

Three consecutive measurements per eye by 2 ophthalmologists with a total of 6 measurements per eye were performed using Lenstar. Axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), and corneal curvatures (K1 and K2) were measured for right and left eyes in 154 school-age children. Right and left eyes were analyzed separately. Each optical parameter obtained by 2 examiners was compared by paired samples t test. Intraobserver and interobserver correlation coefficients (r) were calculated for each variable. Significance was attributed when p<0.05.

RESULTS

Measurements could be performed in 152 of the children included. The mean measurements were not different statistically except for left AL (23.22±0.81 vs 23.21±0.82 mm; p=0.014), left CCT (548.5±32.3 vs 548.0±32.7 µm; p=0.013), and right ACD (3.19±0.25 vs 3.19±0.25 mm; p=0.033). Intraobserver r values were all above 0.957 except for left LT (r=0.786) for examiner 2. Interobserver r values were all above 0.979.

CONCLUSIONS

Although statistical significance was obtained in some variables, they were clinically negligible. The intraobserver and interobserver relation coefficients were very high for all optical parameters including the r value for left LT for examiner 2 (r=0.786). Lenstar is a highly reproducible, user independent, and tolerable instrument for ocular biometry in the pediatric population, with no anesthesia or discomfort.

摘要

目的

评估Lenstar在儿科人群中的可重复性。

方法

两名眼科医生对每只眼睛连续进行三次测量,每只眼睛共进行六次测量,使用Lenstar。对154名学龄儿童的右眼和左眼测量眼轴长度(AL)、中央角膜厚度(CCT)、前房深度(ACD)、晶状体厚度(LT)和角膜曲率(K1和K2)。右眼和左眼分别进行分析。通过配对样本t检验比较两名检查者获得的每个光学参数。计算每个变量的观察者内和观察者间相关系数(r)。当p<0.05时具有统计学意义。

结果

纳入的152名儿童能够进行测量。除左眼AL(23.22±0.81 vs 23.21±0.82 mm;p=0.014)、左眼CCT(548.5±32.3 vs 548.0±32.7 µm;p=0.013)和右眼ACD(3.19±0.25 vs 3.19±0.25 mm;p=0.033)外,平均测量值在统计学上无差异。除检查者2的左眼LT(r=0.786)外,观察者内r值均高于0.957。观察者间r值均高于0.979。

结论

虽然在某些变量中获得了统计学意义,但在临床上可忽略不计。包括检查者2的左眼LT的r值在内,所有光学参数的观察者内和观察者间相关系数都非常高(r=0.786)。Lenstar是一种用于儿科人群眼部生物测量的高度可重复、与用户无关且耐受性良好的仪器,无需麻醉或不适。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验