Central corneal thickness measurements with Fourier-domain optical coherence tomography versus ultrasonic pachymetry and rotating Scheimpflug camera.

机构信息

Department of Ophthalmology, Asahikawa Medical University, Asahikawa, Hokkaido, Japan.

出版信息

Cornea. 2011 Jun;30(6):615-9. doi: 10.1097/ICO.0b013e3181d00800.

Abstract

PURPOSE

To compare the accuracy and repeatability of Fourier-domain optical coherence tomography (FD-OCT) with ultrasonic pachymetry (USP) and a rotating Scheimpflug camera for measuring the central corneal thickness (CCT).

METHODS

The CCT was measured in 30 subjects (30 normal corneas) by the same examiner using RTVue-100 FD-OCT with an anterior segment adaptor, Pentacam rotating Scheimpflug camera, and SP-2000 USP. Two examiners obtained one FD-OCT measurement from 10 eyes of 5 subjects to assess interexaminer reproducibility.

RESULTS

The mean CCT (±SD) measured by FD-OCT, USP, and the Pentacam were 530 ± 33, 544 ± 34, and 552 ± 35 μm, respectively. Significant correlations were found between FD-OCT and USP (r = 0.97; P < 0.0001), FD-OCT and Pentacam (r = 0.97; P < 0.0001), and USP and Pentacam (r = 0.96; P < 0.0001). Pairwise comparisons showed that the FD-OCT CCT measurement was significantly thinner than those of the other 2 methods (P < 0.001 for all comparisons). Regarding intraexaminer repeatability, the intraclass correlation coefficients ranged between 0.97 and 0.98. There was high repeatability of the CCT measurements with all methods. FD-OCT also had high interexaminer reproducibility (intraclass correlation coefficient = 0.98).

CONCLUSIONS

RTVue-100 FD-OCT may be a useful alternative for measuring the CCT; however, it significantly underestimates the CCT compared with the USP and the Pentacam with slight differences. Although highly correlated, the measurements are not directly interchangeable in clinical practice.

摘要

目的

比较傅里叶域光学相干断层扫描(FD-OCT)与超声角膜测厚仪(USP)和旋转式Scheimpflug 相机在测量中央角膜厚度(CCT)方面的准确性和可重复性。

方法

同一位检查者使用配备眼前节适配器的 RTVue-100 FD-OCT、Pentacam 旋转式 Scheimpflug 相机和 SP-2000 USP,对 30 名受试者(30 个正常角膜)的 CCT 进行测量。两名检查者对 5 名受试者的 10 只眼各进行一次 FD-OCT 测量,以评估检查者间的可重复性。

结果

FD-OCT、USP 和 Pentacam 测量的平均 CCT(±SD)分别为 530±33μm、544±34μm 和 552±35μm。FD-OCT 与 USP(r=0.97;P<0.0001)、FD-OCT 与 Pentacam(r=0.97;P<0.0001)和 USP 与 Pentacam(r=0.96;P<0.0001)之间均存在显著相关性。两两比较显示,FD-OCT 的 CCT 测量值明显低于其他 2 种方法(所有比较均 P<0.001)。就检查者内重复性而言,组内相关系数范围在 0.97 至 0.98 之间。所有方法的 CCT 测量值均具有高度重复性。FD-OCT 还具有较高的检查者间可重复性(组内相关系数=0.98)。

结论

RTVue-100 FD-OCT 可能是测量 CCT 的一种有用替代方法;然而,与 USP 和 Pentacam 相比,它显著低估了 CCT,且差异较小。尽管高度相关,但在临床实践中,这些测量值不能直接互换。