Kola Mehmet, Önal Mehmet, Türk Adem, Erdöl Hidayet
Karadeniz Technical University Faculty of Medicine, Department of Ophthalmology, Trabzon, Turkey
Turk J Ophthalmol. 2019 Apr 30;49(2):78-83. doi: 10.4274/tjo.galenos.2018.88972.
To evaluate the repeatability and agreement of macular thickness measurements obtained with E-MM5 and MM6, two different scan modes, on the Optovue RTVue optic coherence tomography (OCT) device.
Three consecutive macular thickness measurements in 30 healthy volunteers were taken using the OCT device E-MM5 and MM6 scan modes. The repeatability and agreement of these measurements obtained from the two scan modes and divided into nine anatomical regions based on early treatment diabetic retinopathy study were subjected to statistical analysis.
The mean age of the participants was 29.7±6.39 years. Intraclass correlation (all ICC values ≥0.86) and coefficient of variation (all coefficient of variation values ≤2%) analyses of consecutive OCT measurements in the nine regions of the macula obtained in both E-MM5 and MM6 scan modes gave high repeatability rates. Mean macular thickness values in the foveal region were 243.76±21.79 μm in E-MM5 mode and 247.04±19.83 μm in MM6 mode (p=0.543). Values for measurements obtained in E-MM5 and MM6 scan modes in parafoveal macular regions were also statistically similar (p>0.05 for all). However, a statistically significant difference was observed between the two modes in perifoveal macular measurements, except in the superior region.
The Optovue RTVue OCT device gives highly repeatable measurement results for macular thicknesses in both E-MM5 and MM6 scan modes. However, it should be considered that measurements performed in E-MM5 and MM6 modes give different results in perifoveal regions.
评估在Optovue RTVue光学相干断层扫描(OCT)设备上使用两种不同扫描模式E-MM5和MM6获得的黄斑厚度测量的重复性和一致性。
使用OCT设备的E-MM5和MM6扫描模式,对30名健康志愿者进行连续三次黄斑厚度测量。根据糖尿病视网膜病变早期治疗研究,将从这两种扫描模式获得的测量结果分为九个解剖区域,并进行重复性和一致性的统计分析。
参与者的平均年龄为29.7±6.39岁。在E-MM5和MM6扫描模式下获得的黄斑九个区域的连续OCT测量的组内相关性(所有ICC值≥0.86)和变异系数(所有变异系数值≤2%)分析显示出高重复性率。在E-MM5模式下,黄斑中心凹区域的平均黄斑厚度值为243.76±21.79μm,在MM6模式下为247.04±19.83μm(p = 0.543)。在黄斑旁中心凹区域,E-MM5和MM6扫描模式下获得的测量值在统计学上也相似(所有p>0.05)。然而,除了上方区域外,在黄斑中心凹周围测量中观察到两种模式之间存在统计学上的显著差异。
Optovue RTVue OCT设备在E-MM5和MM6扫描模式下对黄斑厚度的测量结果具有高度重复性。然而,应考虑到在E-MM5和MM6模式下进行的测量在黄斑中心凹周围区域会给出不同的结果。