Non-contact tonometry in the postoperative eye.

作者信息

Vernon S A

机构信息

Academic Unit of Ophthalmology, University Hospital, Nottingham.

出版信息

Br J Ophthalmol. 1989 Apr;73(4):247-9. doi: 10.1136/bjo.73.4.247.

Abstract

The Keeler Pulsair non-contact tonometer was compared with the Goldmann applanation tonometer in a series of 48 eyes after operation. A correlation coefficient of 0.92 (p less than 0.001) was found between the two instruments, with the Pulsair having a statistically significant tendency to overread the Goldmann slightly in this situation. The Pulsair was, however, shown to be effective in the identification of postoperative ocular hypertension with a high degree of sensitivity and specificity and good patient compliance, while reducing the risk of cross infection.

摘要

在48只术后眼睛中,对Keeler Pulsair非接触眼压计与Goldmann压平眼压计进行了比较。发现两种仪器之间的相关系数为0.92(p小于0.001),在这种情况下,Pulsair眼压计在统计学上有略微高于Goldmann眼压计读数的显著趋势。然而,Pulsair眼压计被证明在识别术后高眼压方面有效,具有高度的敏感性和特异性,患者依从性良好,同时降低了交叉感染的风险。

相似文献

1
Non-contact tonometry in the postoperative eye.术后眼的非接触眼压测量法。
Br J Ophthalmol. 1989 Apr;73(4):247-9. doi: 10.1136/bjo.73.4.247.
5
Evaluation of the Keeler Pulsair non-contact tonometer.基勒脉冲式眼压计的评估。
Acta Ophthalmol (Copenh). 1989 Oct;67(5):567-72. doi: 10.1111/j.1755-3768.1989.tb04109.x.

本文引用的文献

1
Advantages and disadvantages of different methods of senile cataract surgery.
Trans Ophthalmol Soc U K (1962). 1982;102 Pt 3:407-9.
5
Repeated applanation tonometry.多次压平眼压测量法
Am J Ophthalmol. 1968 Jul;66(1):89-91. doi: 10.1016/0002-9394(68)91794-7.
6
Effects of repeated tonometry: genuine and sham measurements.反复眼压测量的效果:真实测量与假测量
Acta Ophthalmol (Copenh). 1972;50(4):574-82. doi: 10.1111/j.1755-3768.1972.tb05987.x.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验