Abstract

BACKGROUND

Technological development and improvements in Wearable Physiological Monitoring devices, have facilitated the wireless and continuous field-based monitoring/capturing of physiologic measures in healthy, clinical or athletic populations. These devices have many applications for prevention and rehabilitation of musculoskeletal disorders, assuming reliable and valid data is collected. The purpose of this study was to appraise the quality and synthesize findings from published studies on psychometric properties of heart rate measurements taken with the Zephyr Bioharness device.

METHODS

We searched the Embase, Medline, PsycInfo, PuMed and Google Scholar databases to identify articles. Articles were appraised for quality using a structured clinical measurement specific appraisal tool. Two raters evaluated the quality and conducted data extraction. We extracted data on the reliability (intra-class correlation coefficients and standard error of measurement) and validity measures (Pearson/Spearman's correlation coefficients) along with mean differences. Agreement parameters were summarised by the average biases and 95% limits of agreement.

RESULTS

A total of ten studies were included: quality ratings ranged from 54 to 92%. The intra-class correlation coefficients reported ranged from 0.85-0.98. The construct validity coefficients compared against gold standard calibrations or other commercially used devices, ranged from 0.74-0.99 and 0.67-0.98 respectively. Zephyr Bioharness agreement error ranged from - 4.81 (under-estimation) to 3.00 (over-estimation) beats per minute, with varying 95% limits of agreement, when compared with gold standard measures.

CONCLUSION

Good to excellent quality evidence from ten studies suggested that the Zephyr Bioharness device can provide reliable and valid measurements of heart rate across multiple contexts, and that it displayed good agreements vs. gold standard comparators - supporting criterion validity.

摘要

背景

可穿戴生理监测设备的技术发展与改进,促进了对健康人群、临床患者或运动员群体生理指标的无线及连续现场监测/采集。若能收集到可靠且有效的数据,这些设备在肌肉骨骼疾病的预防和康复方面有诸多应用。本研究旨在评估已发表的关于使用Zephyr Bioharness设备进行心率测量的心理测量特性研究的质量并综合研究结果。

方法

我们检索了Embase、Medline、PsycInfo、PubMed和谷歌学术数据库以识别文章。使用特定的结构化临床测量评估工具对文章质量进行评估。两名评估者评估质量并进行数据提取。我们提取了关于可靠性(组内相关系数和测量标准误差)和有效性指标(皮尔逊/斯皮尔曼相关系数)以及平均差异的数据。一致性参数通过平均偏差和95%一致性界限进行总结。

结果

共纳入10项研究:质量评分范围为54%至92%。报告的组内相关系数范围为0.85 - 0.98。与金标准校准或其他商业使用设备相比,结构效度系数分别为0.74 - 0.99和0.67 - 0.98。与金标准测量相比,Zephyr Bioharness的一致性误差范围为每分钟 - 4.81(低估)至3.00(高估)次心跳,95%一致性界限各不相同。

结论

来自10项研究的质量证据良好至优秀,表明Zephyr Bioharness设备能够在多种情况下提供可靠且有效的心率测量,并且与金标准比较器相比显示出良好的一致性——支持效标效度。