Suppr超能文献

一次性与可重复使用输尿管肾镜用于逆行性肾内手术(RIRS):三种不同器械的物理和光学性能的系统比较分析。

Single-use versus reusable ureterorenoscopes for retrograde intrarenal surgery (RIRS): systematic comparative analysis of physical and optical properties in three different devices.

机构信息

Department of Urology, Eberhard-Karls-University Tübingen, Hoppe-Seyler-Strasse 3, 72076, Tübingen, Germany.

出版信息

World J Urol. 2018 Dec;36(12):2059-2063. doi: 10.1007/s00345-018-2365-9. Epub 2018 Jun 5.

Abstract

INTRODUCTION

Retrograde intrarenal surgery (RIRS) represents a standard option for kidney stone removal. However, RIRS is considered a cost-intensive procedure. Single-use flexible ureterorenoscopes have been introduced to improve budget predictability in RIRS. We assessed differences in physical and optical properties of single-use devices compared to standard reusable endoscopes.

METHODS

In two single-use (LithoVue™, Boston Scientific; Pusen Uscope UE3011™), and one reusable ureterorenoscope (Flex-Xc™, Karl Storz), we investigated flow rates, deflection, illuminance, and intrapelvic pressure in a porcine kidney model. Subjective image quality was assessed using a standardized questionnaire. Common insertable devices were applied to investigate additional influence on physical properties.

RESULTS

Significant variability in maximum flow rates was observed (Flex-Xc™: 25.8 ml/min, LithoVue™: 30.3 ml/min, Pusen™: 33.4 ml/min, p < 0.05). Insertion of a guide wire resulted in the highest reduction of flow rates in all endoscopes. Flection led to a reduction of absolute flow rates up to 9.4% (Flex-Xc™). Light intensity at 20/50 mm distance was 9090 lx/1857 lx (Flex-Xc™) and 5733 lx/1032 lx (LithoVue™) and 2160 lx/428 lx (Pusen™), respectively (p < 0.05). Subjective image quality score was highest using the Flex-Xc™ endoscope. During manipulation, maximum intrarenal pressure up to 66 mmHg (Pusen™) was measured.

CONCLUSIONS

Significant differences in physical and optical properties of single-use or reusable flexible ureterorenoscopes are present, with putative influence on surgical efficacy and complications. Further comparative evaluation of single-use and reusable endoscopes in a clinical scenario is useful. Moreover, utilization of ureteral access sheaths may be considered to avoid renal damage.

摘要

介绍

逆行肾盂内手术(RIRS)是一种治疗肾结石的标准选择。然而,RIRS 被认为是一种成本密集型的手术。一次性使用的软性输尿管肾镜的出现提高了 RIRS 预算的可预测性。我们评估了一次性使用的设备与标准可重复使用的内窥镜在物理和光学特性方面的差异。

方法

在两个一次性使用的(LithoVue™,波士顿科学公司;Pusen Uscope UE3011™)和一个可重复使用的输尿管肾镜(Flex-Xc™,卡尔史托斯)中,我们在猪肾模型中研究了流量、偏转、照度和肾盂内压力。使用标准化问卷评估主观图像质量。应用常见的插入式设备来研究对物理性能的额外影响。

结果

观察到最大流量率存在显著的可变性(Flex-Xc™:25.8 ml/min,LithoVue™:30.3 ml/min,Pusen™:33.4 ml/min,p<0.05)。插入导丝会导致所有内窥镜中流量率的最大降低。弯曲导致绝对流量率降低最多达 9.4%(Flex-Xc™)。在 20/50 毫米距离处的光强度分别为 9090 lx/1857 lx(Flex-Xc™)和 5733 lx/1032 lx(LithoVue™)和 2160 lx/428 lx(Pusen™)(p<0.05)。使用 Flex-Xc™内窥镜的主观图像质量评分最高。在操作过程中,测量到的最大肾内压力高达 66 毫米汞柱(Pusen™)。

结论

一次性或可重复使用的软性输尿管肾镜在物理和光学特性方面存在显著差异,可能对手术效果和并发症有影响。在临床情况下对一次性使用和可重复使用的内窥镜进行进一步的比较评估是有用的。此外,考虑使用输尿管入路鞘可以避免肾脏损伤。