An Exploratory Study of Washout Rate Analysis for Thallium-201 Single-Photon Emission Computed Tomography Myocardial Perfusion Imaging Using Cadmium Zinc Telluride Detectors.

作者信息

Ishihara Masaru, Onoguchi Masahisa, Shibutani Takayuki

机构信息

1 Department of Radiology, Hyogo Cancer Center, Akashi, Japan.

2 Department of Quantum Medical Technology, Graduate School of Medical Sciences, Kanazawa University, Kanazawa, Japan.

出版信息

Mol Imaging. 2018 Jan-Dec;17:1536012118782322. doi: 10.1177/1536012118782322.

Abstract

The aim of this study was to assess the washout rate (WOR) for thallium-201-chloride single-photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI) using cadmium zinc telluride detectors for SPECT (CZT SPECT) versus conventional Anger-type SPECT (conventional SPECT). A total of 52 Japanese patients were examined using CZT SPECT and conventional SPECT, and the global WORs were compared. Additionally, the MPI WORs were compared for patients with a normal MPI versus those in whom MPI reflected the patients' multivessel disease (MVD) MPI. Washout rates were similar when approximated by CZT SPECT versus conventional SPECT 12.59 ± 2.26%/h vs 12.57 ± 2.27%/h ( P = .997), respectively. The WOR values for CZT SPECT versus conventional SPECT were 13.42%/h (1.53%/h) vs 13.93%/h (1.24%/h) ( P = .337), respectively, for 7 normal MPI patients, and 10.64 ± 2.20%/h vs 10.84 ± 2.26%/h ( P = .848), respectively, for 7 MVD-MPI patients. The WOR values for normal MPI versus MVD-MPI patients for CZT SPECT were 13.42 ± 1.53%/h vs 10.64 ± 2.20%/h ( P = .025), respectively. Thallium-201-chloride WOR values obtained with high-efficiency CZT SPECT, which enabled significantly reduced imaging times and use of a low-dose protocol, were similar to those obtained with conventional SPECT.

摘要

本研究的目的是评估使用碲化镉锌探测器的单光子发射计算机断层扫描(SPECT)(CZT SPECT)与传统安杰型SPECT(传统SPECT)进行氯化铊-201单光子发射计算机断层扫描(SPECT)心肌灌注成像(MPI)的洗脱率(WOR)。共对52名日本患者进行了CZT SPECT和传统SPECT检查,并比较了整体WOR。此外,还比较了MPI正常患者与MPI反映多支血管疾病(MVD)的患者的MPI WOR。CZT SPECT与传统SPECT估算的洗脱率相似,分别为12.59±2.26%/小时和12.57±2.27%/小时(P = 0.997)。对于7名MPI正常的患者,CZT SPECT与传统SPECT的WOR值分别为13.42%/小时(1.53%/小时)和13.93%/小时(1.24%/小时)(P = 0.337);对于7名MVD-MPI患者,分别为10.64±2.20%/小时和10.84±2.26%/小时(P = 0.848)。CZT SPECT的正常MPI患者与MVD-MPI患者的WOR值分别为13.42±1.53%/小时和10.64±2.20%/小时(P = 0.025)。使用高效CZT SPECT获得的氯化铊-201 WOR值与传统SPECT获得的值相似,前者能够显著缩短成像时间并采用低剂量方案。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e410/6077886/e7006d099eaa/10.1177_1536012118782322-fig1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验