Institute of Medical & Biological Engineering, Medical Research Center, Seoul National University, Seoul, Korea.
Interdisciplinary Program in Bioengineering, College of Engineering, Seoul National University, Seoul, Korea.
Sci Rep. 2019 Mar 18;9(1):4747. doi: 10.1038/s41598-019-41066-8.
We propose a noninvasive, self-diagnostic device that enables safe tear collection and glucose measurement. The device described herein was manufactured by tight assembly of a lid for tear collection in conjunction with a strip-type glucose sensor. The lid was designed to be in contact with the inferior palpebral conjunctiva for tear collection and was thus designed to possess a proper contact area and rounded boundaries to avoid eye tissue damage. For the strip-type glucose sensor, we employed a commercially available electrochemical sensor (Accu-Chek test strips), which was modified to reduce the volume of the reaction chamber (0.4 μl) for a small amount of collected tear fluid. When tested with in vivo animal models, the device was able to collect tear fluid in a relatively short time (<2 s) without causing eye tissue damage, and the device allowed the collected tear fluid to be delivered to the sensor for measurement of tear glucose concentrations. The blood glucose concentrations estimated with the tear glucose concentrations obtained with the device exhibited a high correlation with those actually measured with a clinically available glucometer (R = 0.9617).
我们提出了一种无创、自我诊断的设备,可安全地采集泪液并进行葡萄糖测量。本文所述的设备通过紧密组装一个用于收集泪液的盖子和一个条状葡萄糖传感器来制造。盖子的设计与下眼睑结膜接触,以收集泪液,因此设计有适当的接触面积和圆形边界,以避免眼组织损伤。对于条状葡萄糖传感器,我们使用了市售的电化学传感器(Accu-Chek 测试条),并对其进行了修改,以减小反应室的体积(0.4μl),以适应少量采集的泪液。在体内动物模型中进行测试时,该设备能够在相对较短的时间内(<2 秒)收集泪液,而不会造成眼组织损伤,并且该设备允许将收集的泪液输送到传感器以测量泪液中的葡萄糖浓度。该设备测量的泪液葡萄糖浓度与临床可用血糖仪实际测量的血糖浓度之间具有高度相关性(R=0.9617)。