使用IOL Master 500、IOL Master 700和Lenstar LS900进行眼生物测量和屈光结果的比较。

Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900.

作者信息

Song Jae Shin, Yoon Do Yeh, Hyon Joon Young, Jeon Hyun Sun

机构信息

Department of Ophthalmology, Seoul National University College of Medicine, Seoul, Korea.

Department of Ophthalmology, Seoul National University Bundang Hospital, Seongnam, Korea.

出版信息

Korean J Ophthalmol. 2020 Apr;34(2):126-132. doi: 10.3341/kjo.2019.0102.

Abstract

PURPOSE

To evaluate the agreement in ocular biometry outcomes measured by three different devices, the IOL Master 500, IOL Master 700, and Lenstar LS900, and compare the refractive outcomes after cataract surgery obtained using those three devices.

METHODS

Medical records of 178 eyes of 89 patients who underwent ocular biometry with the three devices were retrospectively reviewed, and 124 eyes met the inclusion criteria. Paired comparisons were performed for axial length (AL), mean keratometry (Km), and anterior chamber depth and quantified their agreement using Bland-Altman plots. Subgroup analyses were done according to the AL and the Km. Refractive outcomes were compared with respect to absolute prediction errors after cataract surgery in 54 eyes.

RESULTS

Among 124 eyes, 12, 3, and 5 eyes failed to be measured of AL by IOL Master 500, IOL Master 700 and Lenstar LS900, respectively. The AL measured by Lenstar LS900 was longer than that measured by IOL Master 500 and IOL Master 700 ( < 0.001, = 0.002, respectively). Subgroup analysis revealed that these results were statistically significant only in long eyes (AL >25.5 mm). Km measured using the IOL Master 500 was steeper than that measured with the IOL Master 700 or Lenstar LS900 ( = 0.001, < 0.001, respectively). anterior chamber depth measured by IOL Master 500 was shallower than that measured by IOL Master 700 or Lenstar LS900 ( < 0.001, < 0.001, respectively). Ocular biometry measurements by the three devices showed high agreement with narrow 95% limits of agreement. Absolute prediction errors from the 3 devices showed no statistically significant differences after cataract surgery.

CONCLUSIONS

The IOL Master 700 and Lenstar LS900 demonstrated superior acquisition rates of biometric measurements compared with the IOL Master 500. Ocular biometry using the 3 different devices showed high agreement, although statistically significant differences were observed; however, since there was no difference in predicting the refractive outcomes, those differences are clinically negligible.

摘要

目的

评估三种不同设备IOL Master 500、IOL Master 700和Lenstar LS900测量的眼生物测量结果的一致性,并比较使用这三种设备进行白内障手术后的屈光结果。

方法

回顾性分析89例患者178只眼使用这三种设备进行眼生物测量的病历,124只眼符合纳入标准。对眼轴长度(AL)、平均角膜曲率(Km)和前房深度进行配对比较,并使用Bland-Altman图量化它们的一致性。根据AL和Km进行亚组分析。比较了54只眼白内障手术后的屈光结果与绝对预测误差。

结果

在124只眼中,IOL Master 500、IOL Master 700和Lenstar LS900分别有12只、3只和5只眼未能测量出AL。Lenstar LS900测量的AL比IOL Master 500和IOL Master 700测量的长(分别为<0.001,=0.002)。亚组分析显示,这些结果仅在长眼(AL>25.5mm)中具有统计学意义。使用IOL Master 500测量的Km比使用IOL Master 700或Lenstar LS900测量的更陡(分别为=0.001,<0.001)。IOL Master 500测量的前房深度比IOL Master 700或Lenstar LS900测量的浅(分别为<0.001,<0.001)。三种设备的眼生物测量结果显示出高度一致性,一致性界限狭窄。三种设备的绝对预测误差在白内障手术后无统计学显著差异。

结论

与IOL Master 500相比,IOL Master 700和Lenstar LS900的生物测量获取率更高。使用三种不同设备进行眼生物测量显示出高度一致性,尽管观察到有统计学显著差异;然而,由于在预测屈光结果方面没有差异,这些差异在临床上可以忽略不计。