Sensors (Basel). 2021 Mar 19;21(6):2167. doi: 10.3390/s21062167.
Arterial blood pressure (ABP) is an important vital sign from which it can be extracted valuable information about the subject's health. After studying its morphology it is possible to diagnose cardiovascular diseases such as hypertension, so ABP routine control is recommended. The most common method of controlling ABP is the cuff-based method, from which it is obtained only the systolic and diastolic blood pressure (SBP and DBP, respectively). This paper proposes a cuff-free method to estimate the morphology of the average ABP pulse (ABPM¯) through a deep learning model based on a seq2seq architecture with attention mechanism. It only needs raw photoplethysmogram signals (PPG) from the finger and includes the capacity to integrate both categorical and continuous demographic information (DI). The experiments were performed on more than 1100 subjects from the MIMIC database for which their corresponding age and gender were consulted. Without allowing the use of data from the same subjects to train and test, the mean absolute errors (MAE) were 6.57 ± 0.20 and 14.39 ± 0.42 mmHg for DBP and SBP, respectively. For ABPM¯, R correlation coefficient and the MAE were 0.98 ± 0.001 and 8.89 ± 0.10 mmHg. In summary, this methodology is capable of transforming PPG into an ABP pulse, which obtains better results when DI of the subjects is used, potentially useful in times when wireless devices are becoming more popular.
动脉血压(ABP)是一个重要的生命体征,可以从中提取出有关受试者健康的有价值信息。研究其形态后,可以诊断出高血压等心血管疾病,因此建议定期监测 ABP。控制 ABP 的最常见方法是基于袖带的方法,只能获得收缩压和舒张压(SBP 和 DBP)。本文提出了一种无袖带方法,通过基于 seq2seq 架构的深度学习模型和注意力机制,从手指的原始光体积描记信号(PPG)中估计平均 ABP 脉搏(ABPM¯)的形态。它只需要包含类别和连续人口统计学信息(DI)的能力的原始光电体积描记信号(PPG)。实验在 MIMIC 数据库中进行了 1100 多名受试者,为他们查询了相应的年龄和性别。在不允许使用同一受试者的数据进行训练和测试的情况下,DBP 和 SBP 的平均绝对误差(MAE)分别为 6.57±0.20 和 14.39±0.42mmHg。对于 ABPM¯,R 相关系数和 MAE 分别为 0.98±0.001 和 8.89±0.10mmHg。总之,该方法能够将 PPG 转换为 ABP 脉搏,当使用受试者的 DI 时,它可以获得更好的结果,在无线设备越来越受欢迎的时候,它可能具有一定的实际应用价值。