单容积 H-心脏磁共振波谱技术在 3T 下活体测量心肌肌酸和甘油三酯的比较。

Comparison of single-voxel H-cardiovascular magnetic resonance spectroscopy techniques for in vivo measurement of myocardial creatine and triglycerides at 3T.

机构信息

Aix-Marseille Univ, CNRS, CRMBM UMR 7339, Marseille, France.

APHM, Hôpital Universitaire Timone, CEMEREM, Marseille, France.

出版信息

J Cardiovasc Magn Reson. 2021 May 13;23(1):53. doi: 10.1186/s12968-021-00748-x.

Abstract

BACKGROUND

Single-voxel proton cardiovascular magnetic resonance spectroscopy (H-CMRS) benefits from 3 T to detect metabolic abnormalities with the quantification of intramyocardial fatty acids (FA) and creatine (Cr). Conventional point resolved spectroscopy (PRESS) sequence remains the preferred choice for CMRS, despite its chemical shift displacement error (CSDE) at high field (≥ 3 T). Alternative candidate sequences are the semi-adiabatic Localization by Adiabatic SElective Refocusing (sLASER) recommended for brain and musculoskeletal applications and the localized stimulated echo acquisition mode (STEAM). In this study, we aim to compare these three single-voxel H-CMRS techniques: PRESS, sLASER and STEAM for reproducible quantification of myocardial FA and Cr at 3 T. Sequences are compared both using breath-hold (BH) and free-breathing (FB) acquisitions.

METHODS

CMRS accuracy and theoretical CSDE were verified on a purposely-designed fat-water phantom. FA and Cr CMRS data quality and reliability were evaluated in the interventricular septum of 10 healthy subjects, comparing repeated BH and free-breathing with retrospective gating.

RESULTS

Measured FA/W ratio deviated from expected phantom ratio due to CSDE with all sequences. sLASER supplied the lowest bias (10%, vs -28% and 27% for PRESS and STEAM). In vivo, PRESS provided the highest signal-to-noise ratio (SNR) in FB scans (27.5 for Cr and 103.2 for FA). Nevertheless, a linear regression analysis between the two BH showed a better correlation between myocardial Cr content measured with sLASER compared to PRESS (r = 0.46; p = 0.03 vs. r = 0.35; p = 0.07) and similar slopes of regression lines for FA measurements (r = 0.94; p < 0.001 vs. r = 0.87; p < 0.001). STEAM was unable to perform Cr measurement and was the method with the lowest correlation (r = 0.59; p = 0.07) for FA. No difference was found between measurements done either during BH or FB for Cr, FA and triglycerides using PRESS, sLASER and STEAM.

CONCLUSION

When quantifying myocardial lipids and creatine with CMR proton spectroscopy at 3 T, PRESS provided higher SNR, while sLASER was more reproducible both with single BH and FB scans.

摘要

背景

单体质子心血管磁共振波谱(H-CMRS)受益于 3T 以检测代谢异常,包括心肌内脂肪酸(FA)和肌酸(Cr)的定量。尽管在高场(≥3T)下存在化学位移位移误差(CSDE),但传统的点分辨波谱(PRESS)序列仍然是 CMRS 的首选选择。替代候选序列是半绝热局部选择性反转聚焦(sLASER),推荐用于脑和肌肉骨骼应用,以及局部激发回波采集模式(STEAM)。在这项研究中,我们旨在比较这三种单体 H-CMRS 技术:PRESS、sLASER 和 STEAM,以在 3T 时对心肌 FA 和 Cr 进行可重复的定量。序列使用屏气(BH)和自由呼吸(FB)采集进行比较。

方法

在专门设计的脂肪 - 水仿体上验证了 CMRS 的准确性和理论 CSDE。在 10 名健康受试者的室间隔中评估了 FA 和 Cr CMRS 数据质量和可靠性,比较了重复 BH 和自由呼吸与回顾性门控。

结果

由于所有序列的 CSDE,测量的 FA/W 比与预期的幻影比偏差。sLASER 提供的偏差最小(10%,而 PRESS 和 STEAM 为-28%和 27%)。在体内,PRESS 在 FB 扫描中提供了最高的信噪比(SNR)(Cr 为 27.5,FA 为 103.2)。然而,两种 BH 之间的线性回归分析表明,与 PRESS 相比,sLASER 测量的心肌 Cr 含量具有更好的相关性(r=0.46;p=0.03 与 r=0.35;p=0.07),并且 FA 测量的回归线斜率相似(r=0.94;p<0.001 与 r=0.87;p<0.001)。STEAM 无法进行 Cr 测量,并且对于 FA 的相关性最低(r=0.59;p=0.07)。使用 PRESS、sLASER 和 STEAM,在 BH 或 FB 期间进行 Cr、FA 和甘油三酯测量时,没有发现差异。

结论

在 3T 时使用质子 CMRS 定量心肌脂质和肌酸时,PRESS 提供了更高的 SNR,而 sLASER 则在单次 BH 和 FB 扫描中具有更高的可重复性。