Urias Elizabeth A, Polychronopoulou Efstathia, Pandit Rahul T
Blanton Eye Institute, Houston Methodist Hospital, Houston, TX, USA.
University of Texas Medical Branch, Galveston, TX, USA.
J Ophthalmol. 2021 Sep 16;2021:5860846. doi: 10.1155/2021/5860846. eCollection 2021.
To compare simulated and total keratometry and corneal astigmatism values between the IOLMaster 700 and Galilei G4 devices.
A retrospective chart review was conducted for all patients undergoing phacoemulsification by a single surgeon (RTP) from March through September 2020 and who underwent imaging with both the IOLMaster 700 and Galilei G4. Exclusion criteria were prior corneal surgery, keratectatic diseases and inability to obtain a reliable image during image acquisition. Mean, flat, and steep keratometry values as well as astigmatism magnitude were compared.
A total of 200 eyes of 100 patients were included. Intraclass correlation coefficients (ICC) were moderate or high for all variables. Mean difference ± SD in SimK and TrueK between devices (G4-IOLM) was 0.05 ± 0.318 diopters and -1.1156 ± 0.438 diopters, respectively ( < 0.05 for both). The IOLM measured steeper TrueK value than the G4. For SimK, there was a statistically significant difference between devices only for mean keratometry (K), whereas for TrueK, there were significant differences in flat K, steep K, and mean K. Astigmatism analysis revealed a difference in mean (±SD) SimK of 0.07 (±0.57) D at 94 degrees and in mean TrueK of 0.04 (±0.85) D at 108 degrees.
Though there is overall good correlation between the IOLMaster 700 and Galilei G4 in SimK and astigmatism measurements, there is a significant difference in TrueK measurements, with the IOLM measuring steeper values by about 1.0 diopter as compared to the G4.
比较IOLMaster 700和Galilei G4设备之间模拟角膜曲率和总角膜曲率以及角膜散光值。
对2020年3月至9月由单一外科医生(RTP)进行白内障超声乳化手术且同时使用IOLMaster 700和Galilei G4进行成像的所有患者进行回顾性病历审查。排除标准为既往角膜手术史、角膜扩张性疾病以及在图像采集期间无法获得可靠图像。比较平均、平坦和陡峭角膜曲率值以及散光量。
共纳入100例患者的200只眼。所有变量的组内相关系数(ICC)为中度或高度。设备之间(G4 - IOLM)模拟角膜曲率(SimK)和真实角膜曲率(TrueK)的平均差值±标准差分别为0.05±0.318屈光度和 - 1.1156±0.438屈光度(两者均P<0.05)。IOLMaster 700测量的TrueK值比Galilei G4更陡峭。对于SimK,仅在平均角膜曲率(K)方面设备之间存在统计学显著差异,而对于TrueK,在平坦K、陡峭K和平均K方面均存在显著差异。散光分析显示,在94度时平均(±标准差)SimK的差异为0.07(±0.57)D,在108度时平均TrueK的差异为0.04(±0.85)D。
虽然IOLMaster 700和Galilei G4在SimK和散光测量方面总体相关性良好,但在TrueK测量方面存在显著差异,与Galilei G4相比,IOLMaster 700测量的更陡峭值约高1.0屈光度。