腰椎体间融合中的椎体终板凹陷:锥形 3D 打印多孔钛笼与方形 PEEK 笼。

Vertebral Endplate Concavity in Lateral Lumbar Interbody Fusion: Tapered 3D-Printed Porous Titanium Cage versus Squared PEEK Cage.

机构信息

Department of Orthopedic Surgery, Graduate School of Medicine, Nagoya University, 65 Tsurumai, Showa-ku, Nagoya 466-8560, Japan.

Department of Orthopedic Surgery, Anjo Kosei Hospital, 28 Higashihirokute, Anjo 446-8602, Japan.

出版信息

Medicina (Kaunas). 2023 Feb 15;59(2):372. doi: 10.3390/medicina59020372.

Abstract

: To prevent postoperative problems in extreme lateral interbody fusion (XLIF), it is critical that the vertebral endplate not be injured. Unintentional endplate injuries may depend on the cage. A novel porous titanium cage for XLIF has improved geometry with a tapered tip and smooth surface. We hypothesized that this new cage should lead to fewer endplate injuries. : This retrospective study included 32 patients (mean 74.1 ± 6.7 years, 22 females) who underwent anterior and posterior combined surgery with XLIF for lumbar degenerative disease or adult spinal deformity from January 2018 to June 2022. A tapered 3D porous titanium cage (3DTi; 11 patients) and a squared PEEK cage (sPEEK; 21 patients) were used. Spinal alignment values were measured on X-ray images. Vertebral endplate concavity (VEC) was defined as concavity ≥ 1 mm of the endplate on computed tomography (CT) images, which were evaluated preoperatively and at 1 week and 3 months postoperatively. : There were no significant differences in the patient demographic data and preoperative and 3-month postoperative spinal alignments between the groups. A 3DTi was used for 25 levels and an sPEEK was used for 38 levels. Preoperative local lordotic angles were 4.3° for 3DTi vs. 4.7° for sPEEK ( = 0.90), which were corrected to 12.3° and 9.1° ( = 0.029), respectively. At 3 months postoperatively, the angles were 11.6° for 3DTi and 8.2° for sPEEK ( = 0.013). VEC was present in 2 levels (8.0%) for 3DTi vs. 17 levels (45%) for sPEEK ( = 0.002). After 3 months postoperatively, none of the 3DTi had VEC progression; however, eight (21%) levels in sPEEK showed VEC progression ( = 0.019). : The novel 3DTi cage reduced endplate injuries by reducing the endplate load during cage insertion.

摘要

: 为了防止极外侧椎间融合术(XLIF)术后出现问题,保护椎体终板不受伤至关重要。非故意的终板损伤可能取决于椎间融合器。一种新型的用于 XLIF 的多孔钛笼具有改进的几何形状,尖端逐渐变细,表面光滑。我们假设这种新型笼应能减少终板损伤。 : 这项回顾性研究纳入了 32 名患者(平均年龄 74.1±6.7 岁,22 名女性),他们因腰椎退行性疾病或成人脊柱畸形,于 2018 年 1 月至 2022 年 6 月接受了前路和后路联合 XLIF 手术。使用了锥形 3D 多孔钛笼(3DTi;11 例)和方形聚醚醚酮笼(sPEEK;21 例)。在 X 射线图像上测量脊柱对线值。在 CT 图像上,定义椎体终板凹陷(VEC)为终板凹陷≥1mm,分别在术前和术后 1 周及 3 个月进行评估。 : 两组患者的人口统计学数据和术前及术后 3 个月的脊柱对线无显著差异。3DTi 用于 25 个节段,sPEEK 用于 38 个节段。3DTi 的术前局部前凸角为 4.3°,sPEEK 为 4.7°( = 0.90),术后分别矫正至 12.3°和 9.1°( = 0.029)。术后 3 个月,3DTi 为 11.6°,sPEEK 为 8.2°( = 0.013)。3DTi 有 2 个节段(8.0%)出现 VEC( = 0.002),sPEEK 有 17 个节段(45%)出现 VEC( = 0.002)。术后 3 个月,3DTi 无 VEC 进展,而 sPEEK 有 8 个(21%)节段出现 VEC 进展( = 0.019)。 : 新型 3DTi 笼通过减少笼插入过程中对终板的负荷,减少了终板损伤。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/c7c9/9967078/910985d343f1/medicina-59-00372-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验