Wang W L, Blaser M J
J Clin Microbiol. 1986 Apr;23(4):709-14. doi: 10.1128/jcm.23.4.709-714.1986.
Because differences in recognition of Campylobacter fetus and Campylobacter jejuni in systemic infections may be due partially to differences in the ability to cultivate these organisms, we studied their growth characteristics in two widely used blood culture systems. In the Roche Septi-Chek system (Hoffman-La Roche, Inc., Nutley, N.J.), over a broad range of inocula all strains were detected in broth within 2 days and on paddles within 3 days. In the BACTEC 6B aerobic bottles (Johnston Laboratories, Inc., Towson, Md.), C. jejuni and C. fetus took a median of 5 and 3 days, respectively, to reach the growth index threshold. However, in the BACTEC 7D anaerobic bottles, C. fetus required a median of 2 days to reach the growth index threshold, whereas for C. jejuni the median was greater than 10 days. The poor performance of C. jejuni in both BACTEC systems may have been due to unfavorable incubation atmospheres and may partially explain why C. jejuni bacteremia is so infrequently detected. Overall, the Roche Septi-Chek system was excellent for detecting Campylobacter strains in blood cultures.
由于全身性感染中胎儿弯曲杆菌和空肠弯曲杆菌的识别差异可能部分归因于培养这些微生物的能力差异,我们研究了它们在两种广泛使用的血培养系统中的生长特性。在罗氏Septi-Chek系统(新泽西州纳特利的霍夫曼-罗氏公司)中,在广泛的接种范围内,所有菌株在2天内在肉汤中被检测到,在3天内在培养板上被检测到。在BACTEC 6B需氧瓶(马里兰州陶森的约翰斯顿实验室公司)中,空肠弯曲杆菌和胎儿弯曲杆菌分别中位数需要5天和3天达到生长指数阈值。然而,在BACTEC 7D厌氧瓶中,胎儿弯曲杆菌中位数需要2天达到生长指数阈值,而空肠弯曲杆菌中位数大于10天。空肠弯曲杆菌在两种BACTEC系统中的表现不佳可能是由于培养环境不利,这可能部分解释了为何空肠弯曲杆菌菌血症很少被检测到。总体而言,罗氏Septi-Chek系统在检测血培养中的弯曲杆菌菌株方面表现出色。