Kannan Mageshraja, Saminathan Sathiyan, Chandraraj Varatharaj, Gowtham Raj D, Ganesh K M
Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Bengaluru, Karnataka, India.
Rep Pract Oncol Radiother. 2023 Jun 26;28(2):241-254. doi: 10.5603/RPOR.a2023.0024. eCollection 2023.
Beam matching is widely used to ensure that linear accelerators used in radiotherapy have equal dosimetry characteristics. Small-field output factors (OF) were measured using different detectors infour beam-matched linear accelerators and the measured OFs were compared with existing treatment planning system (TPS) Monte Carlo algorithm calculated OFs.
Three Elekta Versa HD and one Elekta Infinitylinear accelerators with photon energies of 6 MV flattening filter (FF), 10 MVFF, 6 MV flattening filter free (FFF) and 10 MVFFF were used in this study. All the Linac'swere beam-matched, Dosimetry beam data were ± 1% compare with Reference Linac. Ten different type of detectors (four ionizationchambers and six diode detectors) were used for small-field OF measurements. The OFs were measured for field sizes of 1 × 1 to 10 × 10 cm, and normalized to 10 × 10 cm field size. The uncorrected and corrected OFs were calculated from these measurements. The corrected OF was compare with existing treatment planning system (TPS) Monte Carlo algorithm calculated OFs.
The small-field corrected and Uncorrected OF variations among the linear accelerators was within 1% for all energies and detectors. An increase in field size led to a reduction in the difference between OFs among the detectors, which was the case for all energies. The RSD values decreased with increasing field size. The TRS 483 provided Detector-specificoutput-correction factor (OCF) reduced uncertainty in small-field measurements.
It is necessary to implement the OF-correction of small fields in a TPS. Special care must be taken to incorporate the corrected small-field OF in a TPS.
射野匹配广泛用于确保放射治疗中使用的直线加速器具有相同的剂量学特征。使用不同探测器测量了四台射野匹配直线加速器的小射野输出因子(OF),并将测量得到的OF与现有治疗计划系统(TPS)蒙特卡罗算法计算得到的OF进行比较。
本研究使用了三台医科达Versa HD直线加速器和一台医科达Infinity直线加速器,光子能量分别为6 MV均整滤过器(FF)、10 MV FF、6 MV无均整滤过器(FFF)和10 MV FFF。所有直线加速器均进行了射野匹配,剂量学射野数据与参考直线加速器相比在±1%以内。使用十种不同类型的探测器(四个电离室和六个二极管探测器)进行小射野OF测量。测量了1×1至10×10 cm射野大小的OF,并将其归一化为10×10 cm射野大小。根据这些测量计算未校正和校正后的OF。将校正后的OF与现有治疗计划系统(TPS)蒙特卡罗算法计算得到的OF进行比较。
对于所有能量和探测器,直线加速器之间小射野校正和未校正的OF变化在1%以内。射野大小增加导致探测器之间OF差异减小,所有能量均如此。相对标准偏差(RSD)值随射野大小增加而降低。TRS 483提供的特定探测器输出校正因子(OCF)降低了小射野测量中的不确定性。
在TPS中实施小射野的OF校正很有必要。必须特别注意将校正后的小射野OF纳入TPS。