Sardari Sara, Fotouhi Akbar, Jafarzadehpur Ebrahim, Khabazkhoob Mehdi
Research and Technology Deputy, Tehran University of Medical Sciences, Tehran 1417653761, Iran.
Noor Ophthalmology Research Center, Noor Eye Hospital, Tehran 1968653111, Iran.
Int J Ophthalmol. 2023 Dec 18;16(12):2095-2104. doi: 10.18240/ijo.2023.12.24. eCollection 2023.
To determine the agreement of ocular biometric indices including axial length, keratometric readings, anterior chamber depth, and horizontal corneal diameter between the Pentacam AXL and IOL Master 500.
The study was a large cross-sectional population-based study (Tehran Geriatric Eye Study) conducted from Jan 2019 to Jan 2020. A total of 160 clusters were randomly selected proportional to size (each cluster contained 20 individuals) from 22 strata of Tehran city. All people aged 60y and above were invited to participate in the study. For all participants, preliminary ocular examinations were performed including the measurement of uncorrected and best-corrected visual acuity, objective and subjective refraction, anterior and posterior segment examinations. All participants underwent an ocular biometry using the Pentacam AXL and IOL Master 500.
The 95% limits of agreement (LoA) between the two devices were -0.13 to 0.19, -0.15 to 0.17, and -0.13 to 0.19 in normal, pseudophakic, and cataractous eyes, respectively. With increasing the axial length, the difference between the two devices significantly increased in all three groups of normal, pseudophakic, and cataractous eyes (<0.001). The 95% LoAs between the two devices regarding the mean keratometry shows that the best LoAs were seen in cataractous (-0.33 to 0.81) and followed by normal eyes (-0.36 to 0.86) and the pseudophakic eyes (-0.48 to 0.90) had the widest LoA. The 95% LoAs for horizontal corneal diameter measurements were -0.08 to 0.86, -0.03 to 0.83, and -0.07 to 0.87 in normal, pseudophakic, and cataractous eyes, respectively. The 95% LoAs of anterior chamber depth measurements between the two devices was -0.39 to 0.19 and -0.37 to 0.13 in normal eyes and cataractous, respectively.
The Pentacam AXL has excellent agreement with the gold standard, IOL Master 500 in measuring axial length. In eyes with cataracts, the difference between the two devices is more scattered. With the increasing of axial length, the difference between the two devices increased, which should be considered when using Pentacam AXL.
确定Pentacam AXL与IOL Master 500之间眼生物测量指标的一致性,这些指标包括眼轴长度、角膜曲率读数、前房深度和水平角膜直径。
该研究是一项基于人群的大型横断面研究(德黑兰老年眼病研究),于2019年1月至2020年1月进行。从德黑兰市的22个阶层中按比例随机抽取160个聚类(每个聚类包含20人)。邀请所有60岁及以上的人参与研究。对所有参与者进行了初步眼科检查,包括测量未矫正和最佳矫正视力、客观和主观验光、眼前段和眼后段检查。所有参与者均使用Pentacam AXL和IOL Master 500进行了眼生物测量。
在正常眼、人工晶状体眼和白内障眼中,两种设备之间的95%一致性界限(LoA)分别为-0.13至0.19、-0.15至0.17和-0.13至0.19。随着眼轴长度增加,在正常眼、人工晶状体眼和白内障眼这三组中,两种设备之间的差异均显著增加(<0.001)。两种设备关于平均角膜曲率的95% LoA表明,白内障眼中的最佳LoA为(-0.33至0.81),其次是正常眼(-0.36至0.86),人工晶状体眼的LoA最宽(-0.48至0.90)。水平角膜直径测量的95% LoA在正常眼、人工晶状体眼和白内障眼中分别为-0.08至0.86、-0.03至0.83和-0.07至0.87。两种设备在前房深度测量方面的95% LoA在正常眼中为-0.39至0.19,在白内障眼中为-0.37至0.13。
Pentacam AXL在测量眼轴长度方面与金标准IOL Master 500具有良好的一致性。在白内障眼中,两种设备之间的差异更分散。随着眼轴长度增加,两种设备之间的差异增大,使用Pentacam AXL时应予以考虑。