Brannon P, Kiehn T E
J Clin Microbiol. 1985 Dec;22(6):951-4. doi: 10.1128/jcm.22.6.951-954.1985.
The Isolator 10 lysis-centrifugation blood culture system (E. I. du Pont de Nemours & Co., Inc., Wilmington, Del.) was compared with the BACTEC radiometric method (Johnston Laboratories, Inc., Towson, Md.) with 6B and 7D broth media for the recovery of bacteria and yeasts. From 11,000 blood cultures, 1,174 clinically significant organisms were isolated. The Isolator system recovered significantly more total organisms, members of the family Enterobacteriaceae, Staphylococcus spp., and yeasts. The BACTEC system recovered significantly more Pseudomonas spp., Streptococcus spp., and anaerobes. Of the Isolator colony counts, 87% measured less than 11 CFU/ml of blood. Organisms, on an average, were detected the same day from each of the two culture systems. Only 13 of the 975 BACTEC isolates (0.01%) were recovered by subculture of growth-index-negative bottles, and 12 of the 13 were detected in another broth blood culture taken within 24 h. Contaminants were recovered from 4.8% of the Isolator 10 and 2.3% of the BACTEC cultures.
将10升隔离器裂解离心血液培养系统(E.I.杜邦公司,特拉华州威尔明顿)与BACTEC放射性测量法(约翰斯顿实验室公司,马里兰州陶森),采用6B和7D肉汤培养基,用于细菌和酵母菌的回收培养,并进行了比较。在11000份血液培养样本中,分离出1174株具有临床意义的微生物。隔离器系统回收的微生物总数、肠杆菌科成员、葡萄球菌属和酵母菌明显更多。BACTEC系统回收的假单胞菌属、链球菌属和厌氧菌明显更多。在隔离器菌落计数中,87%的计数低于每毫升血液11个菌落形成单位。两种培养系统平均在同一天检测到微生物。975份BACTEC分离株中只有13株(0.01%)通过生长指数阴性瓶的传代培养回收,其中13株中的12株在24小时内采集的另一管肉汤血液培养中检测到。从4.8%的隔离器10培养样本和2.3%的BACTEC培养样本中回收了污染物。