Lampe A S, van der Reijden T J
Eur J Clin Microbiol. 1984 Aug;3(4):301-5. doi: 10.1007/BF01977477.
Three commercially available test systems for the identification of nonfermentative gram-negative rods were compared: API 20NE, flow N/F and Minitek Nonfermenter System. Two hundred strains were identified by conventional means and by each test system. The rate of correct identification of Acinetobacter, Alcaligenes, Flavobacterium and Moraxella strains to genus level and of the other genera to species level was 92% with API 20NE, 84% with flow N/F and 75% with Minitek Nonfermenter System. The need for these kits in the diagnostic hospital laboratory is also discussed.
API 20NE、flow N/F和Minitek非发酵菌系统。通过传统方法和每个检测系统对200株菌株进行了鉴定。API 20NE对不动杆菌属、产碱杆菌属、黄杆菌属和莫拉菌属菌株到属水平以及对其他属到种水平的正确鉴定率为92%,flow N/F为84%,Minitek非发酵菌系统为75%。还讨论了这些试剂盒在医院诊断实验室中的必要性。