Krisher K K, Whyburn D R, Koepnick F E
Clinical Microbiology Laboratory, Children's Medical Center of Dallas, Texas 75235.
J Clin Microbiol. 1993 Apr;31(4):793-7. doi: 10.1128/jcm.31.4.793-797.1993.
The performance of the Pedi-BacT system, the BacT/Alert (Organon Teknika Corp., Durham, N.C.) pediatric blood culture bottle, was compared with that of a conventional 20-ml supplemented peptone broth tube (Becton-Dickinson Corp., Cockeysville, Md.) (BD system) in matched aerobic cultures. The tubes of the BD system were visually examined daily for 7 days and were subcultured during the first 24 h of incubation. Pedi-BacT cultures were mechanically agitated and continuously monitored for growth by the instrument. Of the 6,628 compliant pairs, 331 (5.0%) were positive in both systems, 220 (3.3%) were positive in the Pedi-BacT system only, and 170 (2.6%) were positive in the BD system only. One (0.02%) false-negative culture and 15 (0.2%) false-positive cultures occurred with the Pedi-BacT system while 20 (0.3%) false-negative cultures and 35 (0.5%) false-positive cultures occurred with the BD system. Of 288 clinically significant organisms detected in matched pairs from which a single isolate was recovered, 176 (61%) were recovered from both systems, 83 (29%) were recovered from the Pedi-BacT system only (P < 0.0001), and 29 (10%) were recovered from the BD system only. Members of the family Enterobacteriaceae (P < 0.01), miscellaneous nonfermenters (P < 0.05), and Candida spp. (P < 0.01) were isolated more frequently in the Pedi-BacT system than in the BD system. No significant difference in recovery of other organisms was found between the systems. The average time to detection for the Pedi-BacT system ranged from 11.5 h for streptococci to 29.7 h for enterococci, while that for the BD system ranged from 20.3 h for streptococci to 66.4 h for some nonfermenters. The BacT/Alert system is a reliable, labor-saving alternative to conventional blood culture methods.
在配对的需氧培养中,对Pedi - BacT系统(BacT/Alert儿科血培养瓶,Organon Teknika公司,北卡罗来纳州达勒姆)与传统的20毫升补充蛋白胨肉汤管(Becton - Dickinson公司,马里兰州科基斯维尔)(BD系统)的性能进行了比较。BD系统的试管每天进行肉眼检查,持续7天,并在培养的前24小时进行传代培养。Pedi - BacT培养物进行机械搅拌,并由仪器持续监测生长情况。在6628对符合要求的配对培养中,两个系统均呈阳性的有331对(5.0%),仅Pedi - BacT系统呈阳性的有220对(3.3%),仅BD系统呈阳性的有170对(2.6%)。Pedi - BacT系统出现1例(0.02%)假阴性培养和15例(0.2%)假阳性培养,而BD系统出现20例(0.3%)假阴性培养和35例(0.5%)假阳性培养。在从配对培养中检测到的288种临床意义重大的微生物中,从单个分离物中回收的,两个系统均回收的有176种(61%),仅从Pedi - BacT系统回收的有83种(29%)(P < 0.0001),仅从BD系统回收的有29种(10%)。肠杆菌科成员(P < 0.01)、其他非发酵菌(P < 0.05)和念珠菌属(P < 0.01)在Pedi - BacT系统中的分离频率高于BD系统。两个系统在其他微生物的回收方面未发现显著差异。Pedi - BacT系统的平均检测时间从链球菌的11.5小时到肠球菌的29.7小时不等,而BD系统的平均检测时间从链球菌的20.3小时到一些非发酵菌的66.4小时不等。BacT/Alert系统是传统血培养方法的一种可靠、省力的替代方法。