The validity of the telemetric system CORTEX X1 in the ventilatory and gas exchange measurement during exercise.

作者信息

Schulz H, Helle S, Heck H

机构信息

Department of Sportsmedicine, Ruhr-University Bochum, Germany.

出版信息

Int J Sports Med. 1997 Aug;18(6):454-7. doi: 10.1055/s-2007-972663.

Abstract

With the portable spirograph CORTEX X1 both the oxygen consumption and the carbon dioxide output can be determined. Therefore, the aim of the present study was to determine the accuracy of the CORTEX X1 in measuring F(E)O2, F(E)CO2 and V(E) when attached to a motor-driven mechanical syringe and to validate the CORTEX X1 against a standardized breath-by-breath system during a graded bicycle ergometry. Fifteen subjects (8 male, 7 female; 26.7+/-3.3 years) performed two graded exercise tests on a bicycle ergometer (50 W incline every 3 min) until volitional fatigue in randomized order. During rest and during the last 30 s of each step ventilatory and gas exchange parameters were measured with the CORTEX X1 and the OXYCONgamma. At rest and at each step no significant differences exist for VO2 (F = 0.97) and VCO2 (F = 0.90). The orthogonal regression equation of the VO2-values was VO2(X1) = -75.5 + 1.01 x VO2(Oxy) and the equation of the VCO2-values was VCO2(X1) = 21.7-1.008 x VCO2(Oxy). The VO2-max-values were nearly the same: 3569+/-924 ml/min (Oxy) and 3497+/-993ml/min (X1). Similar findings were made with regard to VCO2max 4117+/-1010 ml/min (Oxy) and 4126+/-1090 ml/ min. (X1). Maximal values for heart rate were 181+/-10 beats/ min (X1) and 180+/-8 beats/min (Oxy) (F=0.21), for maximal power 256+/-64 W (X1) and 257+/-63 (Oxy) (F = 0.0001) and for maximal ventilation 118+/-31 l/min) and 120+/-35 l/min (Oxy) (F = 0.03) with no significant difference. When attached to the motor-driven syringe V(E) was accurately measured up to 288 l/ min. Over a period of 40 min there was no drift observed in F(E)O2 and F(E)CO2. In conclusion, with the CORTEX X1 VO2 and VCO2 can be accurately determined.

摘要

使用便携式肺量计CORTEX X1可以测定耗氧量和二氧化碳排出量。因此,本研究的目的是确定连接到电动机械注射器时,CORTEX X1在测量呼出氧分数(F(E)O2)、呼出二氧化碳分数(F(E)CO2)和每分通气量(V(E))方面的准确性,并在分级自行车测力计运动期间,与标准化逐次呼吸系统对比验证CORTEX X1。15名受试者(8名男性,7名女性;年龄26.7±3.3岁)以随机顺序在自行车测力计上进行两次分级运动测试(每3分钟递增50W),直至自愿疲劳。在休息期间以及每一级运动的最后30秒,使用CORTEX X1和OXYCONgamma测量通气和气体交换参数。在休息时以及每一级运动时,耗氧量(F = 0.97)和二氧化碳排出量(F = 0.90)均无显著差异。耗氧量值的正交回归方程为VO2(X1) = -75.5 + 1.01×VO2(Oxy),二氧化碳排出量值的方程为VCO2(X1) = 21.7 - 1.008×VCO2(Oxy)。最大耗氧量值几乎相同:3569±924毫升/分钟(Oxy)和3497±993毫升/分钟(X1)。关于最大二氧化碳排出量也有类似发现:4117±1010毫升/分钟(Oxy)和4126±1090毫升/分钟(X1)。心率最大值分别为181±10次/分钟(X1)和180±8次/分钟(Oxy)(F = 0.21),最大功率分别为256±64W(X1)和257±63W(Oxy)(F = 0.0001),最大通气量分别为118±31升/分钟和120±35升/分钟(Oxy)(F = 0.03),均无显著差异。连接到电动注射器时,每分通气量(V(E))在高达288升/分钟时可被准确测量。在40分钟的时间段内,呼出氧分数(F(E)O2)和呼出二氧化碳分数(F(E)CO2)未观察到漂移。总之,使用CORTEX X1可以准确测定耗氧量和二氧化碳排出量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验