Suppr超能文献

血液透析期间通路血流量测量中稀释技术的准确性。

Accuracy of dilution techniques for access flow measurement during hemodialysis.

作者信息

Krivitski N M, MacGibbon D, Gleed R D, Dobson A

机构信息

Transonic Systems Inc, Ithaca, NY 14850, USA.

出版信息

Am J Kidney Dis. 1998 Mar;31(3):502-8. doi: 10.1053/ajkd.1998.v31.pm9506688.

Abstract

Access flow is now widely measured by creating artificial recirculation with the dialysis lines reversed and using dilution methods that sense either ultrasound velocity, electrical impedance, optical, or thermal changes. This study identifies and quantifies factors that influence the accuracy of access flow measurements and recommends ways to reduce these errors. Two major sources of access flow measurement error are identified, arising firstly from the second pass of the indicator by recirculation through the cardiopulmonary system (cardiopulmonary recirculation, CPR), and secondly from changes in venous line blood flow (Qb) and vascular access flow induced by the pressure of venous bolus injections. These errors are considered from theory, by direct measurement of access flow in a sheep model, and by analysis of clinical data. Two extremes for the venous introduction of indicator can be considered in access flow measurements, a slow infusion, which perturbs neither the venous line flow nor access flow but increases the error attributable to the second pass of the indicator by recirculation through cardiopulmonary system, or rapid injection, which eases separation of the second pass of the indicator signal but generates changes in the venous flow and access flow. If CPR is not eliminated, the area added to that of the first pass of indicator ranges up to 40%. Good time resolution could permit the separation of the areas generated by the first and second passage of the indicator. In sheep experiments, injections of 5 or 10 mL into a venous port close to the vascular access caused Qb to change by 20% to 40%. Both the animal experiments and analysis of raw data collected during routine clinical dialysis showed that moving the injection site sufficiently far from the patient, before or into the venous bubble trap, reduced the increase in Qb to only approximately 5% during the critical time when the concentration curve is changing for most tubing brands (Baxter, Belco, Gambro, Hospal, Medisystem, and National Medical Care). Because of the smaller volume of the venous bubble chamber in Cobe tubing (Cobe, Centrysystem 3), this brand showed approximately a 20% increase in Qb. Moving the site of bolus injections to before the bubble trap in the sheep experiments also eliminated the influence of changes in access flow. An additional error in access flow measurement of 20% or more arises from the use of flow reading taken from pump setting rather than a measured flow. The discrepancy between the real flow and pump setting is attributable to needle size, vascular access conditions, or pump calibration. The results show that problems can be minimized by using a dual sensor system that retains the precise timing necessary for separation of access recirculation from CPR; by accurate measurement of dialyzer blood flow; by moving the site of injection to before the venous bubble trap, sufficiently far from the patient, and correcting for any remaining deviations in flow in the venous line concurrent with the dilution curve.

摘要

目前,通过反转透析管路创建人工再循环,并使用检测超声速度、电阻抗、光学或热变化的稀释方法来广泛测量通路血流量。本研究识别并量化了影响通路血流量测量准确性的因素,并推荐了减少这些误差的方法。确定了通路血流量测量误差的两个主要来源,首先是指示剂通过心肺系统再循环的第二次通过(心肺再循环,CPR),其次是静脉推注压力引起的静脉管路血流量(Qb)和血管通路血流量的变化。通过理论分析、在绵羊模型中直接测量通路血流量以及分析临床数据来考虑这些误差。在通路血流量测量中,可以考虑指示剂静脉引入的两个极端情况,一种是缓慢输注,它既不会干扰静脉管路血流量也不会干扰通路血流量,但会增加由于指示剂通过心肺系统再循环的第二次通过而导致的误差,另一种是快速注射,它便于分离指示剂信号的第二次通过,但会产生静脉血流量和通路血流量的变化。如果不消除CPR,添加到指示剂第一次通过面积上的面积范围可达40%。良好的时间分辨率可以允许分离指示剂第一次和第二次通过产生的面积。在绵羊实验中,向靠近血管通路的静脉端口注射5或10 mL会导致Qb变化20%至40%。动物实验和常规临床透析期间收集的原始数据分析均表明,在大多数管路品牌(百特、贝尔科、金宝、博朗、美迪生和国家医疗保健)浓度曲线变化的关键时间内,将注射部位移至离患者足够远的位置,即在静脉气泡捕集器之前或进入静脉气泡捕集器,可将Qb的增加降低至仅约5%。由于科倍管路(科倍,Centrysystem 3)中的静脉气泡腔体积较小,该品牌的Qb增加了约20%。在绵羊实验中将推注注射部位移至气泡捕集器之前也消除了通路血流量变化的影响。通路血流量测量中另外20%或更多的误差来自于使用从泵设置读取的流量读数而非测量的流量。实际流量与泵设置之间的差异归因于针头尺寸、血管通路状况或泵校准。结果表明,通过使用双传感器系统可以将问题最小化,该系统保留了从CPR中分离通路再循环所需的精确时间;通过准确测量透析器血流量;通过将注射部位移至离患者足够远的静脉气泡捕集器之前,并校正静脉管路中与稀释曲线同时出现的任何剩余流量偏差。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验