Paul Tania, Lim Mira, Starr Christopher E, Lloyd Harriet O, Coleman D Jackson, Silverman Ronald H
Department of Ophthalmology, Weill Cornell Medical College, New York Presbyterian Hospital, New York, New York 10065, USA.
J Cataract Refract Surg. 2008 Nov;34(11):1906-12. doi: 10.1016/j.jcrs.2008.07.013.
To compare central corneal thickness (CCT) measurements by the Orbscan II device, contact ultrasound (US) pachymetry, and the noncontact Artemis 2 scanning US system.
Department of Ophthalmology, Weill Cornell Medical College, New York, New York, USA.
The CCT in 40 eyes (20 normal subjects) was measured by the Orbscan II followed by contact US pachymetry and then the Artemis 2. Results were compared using analysis of variance (ANOVA), paired t tests, and Bland-Altman plots.
There was a significant difference in CCT measurements between the 3 modes (F = 32.84, P = .0001, 1-way ANOVA). Artemis 2 and US pachymetry measurements were highly correlated (r2 = 0.963, P < .0001), although Artemis 2 values were a mean of 11.2 microm +/- 6.6 (SD) thinner than pachymetry values. Artemis 2 and Orbscan II measurements were less well correlated (r2 = 0.851, P < .001); Orbscan II values were a mean of 7.5 +/- 15.7 microm thinner than Artemis 2 values. Orbscan II values showed a trend toward increasing underestimation of CCT in thinner corneas.
Ultrasound pachymetry and Artemis 2 CCT measurements were highly correlated; the 11 microm mean difference in measurements may be attributed to decentration, oblique incidence of the probe to the cornea, or possibly the effect of topical anesthesia with contact pachymetry. Although the mean difference between Orbscan II and Artemis 2 values was 7.5 microm, Orbscan values were less correlated than Artemis 2 values with contact US pachymetry and were prone to underestimation of the CCT in thinner corneas.
比较Orbscan II设备、接触式超声测厚法及非接触式Artemis 2扫描超声系统测量中央角膜厚度(CCT)的结果。
美国纽约州纽约市威尔康乃尔医学院眼科。
对40只眼(20名正常受试者)先使用Orbscan II测量CCT,接着采用接触式超声测厚法,然后使用Artemis 2进行测量。使用方差分析(ANOVA)、配对t检验及Bland-Altman图对结果进行比较。
三种测量模式间的CCT测量结果存在显著差异(F = 32.84,P = 0.0001,单因素方差分析)。Artemis 2与超声测厚法的测量结果高度相关(r2 = 0.963,P < 0.0001),不过Artemis 2测量值平均比测厚法测量值薄11.2微米±6.6(标准差)。Artemis 2与Orbscan II测量结果的相关性较差(r2 = 0.851,P < 0.001);Orbscan II测量值平均比Artemis 2测量值薄7.5±15.7微米。在较薄角膜中,Orbscan II测量值显示出对CCT低估增加的趋势。
超声测厚法与Artemis 2测量的CCT高度相关;测量值平均相差11微米可能归因于偏心、探头与角膜的斜入射,或接触式测厚法中表面麻醉的影响。虽然Orbscan II与Artemis 2测量值的平均差异为7.5微米,但Orbscan测量值与接触式超声测厚法的相关性低于Artemis 2测量值,且在较薄角膜中容易低估CCT。