Suppr超能文献

便携式流量容积仪Microspiro HI-298的准确性、精密度和线性度。

Accuracy, precision and linearity of the portable flow-volume meter Microspiro HI-298.

作者信息

Dompeling E, van Schayck C P, Folgering H, van den Hoogen H J, van Weel C

机构信息

Dept of General Practice/Family Medicine, Nijmegen University, The Netherlands.

出版信息

Eur Respir J. 1991 May;4(5):612-5.

PMID:1936232
Abstract

The accuracy, precision and linearity of a new portable flow-volume meter, the Microspiro HI-298 (Chest Corporation, Tokyo, Japan), was investigated using a Fleisch no. 4 pneumotachograph as a standard. After connection and calibration of the pneumotachograph and the Microspiro, a healthy subject performed 44 forced vital capacity (FVC) manoeuvres at different levels of lung inflation. The FVC of these expirations ranged from 2.5-5.1 l. Linear regression of Microspiro values (dependent variable) on Fleisch pneumotachograph values (independent variable) showed that a good linear relationship existed: Pearson correlation coefficients ranged from 0.938-0.985. Linearity of the Microspiro was good except for the peak expiratory flow rate (PEFR) and the maximal expiratory flow at 25% of the expired volume (MEF75). The random error (measure of precision) of all flow-volume (F-V) indices was lower than 5%. The systematic error (measure of accuracy) was low for the forced expiratory volume in one second (FEV1) and the FVC (1% and 4.6%, respectively) but much higher for the instantaneous expiratory flows (PEFR 11.0%; MEF75 7.0%; MEF50 8.5%; MEF25 11.4%). Only the total error in FEV1 complied with the tolerance of 4% of the European Community for Coal and Steel (ECCS). When the measured values were adjusted according to the regression equations of this study, all F-V indices were accurate and precise within 5%. It was concluded that the portable Microspiro HI-298 is a useful instrument for bedside, work-site spirometry and for use in general practice.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

以 Fleisch 4 号呼吸流速仪作为标准,对一款新型便携式流量 - 容积仪 Microspiro HI - 298(日本东京 Chest 公司)的准确性、精密度和线性进行了研究。在连接并校准呼吸流速仪和 Microspiro 后,一名健康受试者在不同肺充气水平下进行了 44 次用力肺活量(FVC)动作。这些呼气的 FVC 范围为 2.5 - 5.1 升。Microspiro 值(因变量)对 Fleisch 呼吸流速仪值(自变量)的线性回归表明存在良好的线性关系:Pearson 相关系数范围为 0.938 - 0.985。除了呼气峰值流速(PEFR)和呼出容积 25%时的最大呼气流量(MEF75)外,Microspiro 的线性良好。所有流量 - 容积(F - V)指标的随机误差(精密度度量)低于 5%。一秒用力呼气容积(FEV1)和 FVC 的系统误差(准确性度量)较低(分别为 1%和 4.6%),但瞬时呼气流量的系统误差要高得多(PEFR 为 11.0%;MEF75 为 7.0%;MEF50 为 8.5%;MEF25 为 11.4%)。只有 FEV1 的总误差符合欧洲煤钢共同体(ECCS)4%的公差要求。当根据本研究的回归方程调整测量值时,所有 F - V 指标在 5%范围内准确且精密。得出结论,便携式 Microspiro HI - 298 是一种用于床边、工作场所肺活量测定以及一般临床实践的有用仪器。(摘要截取自 250 字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验