Transcutaneous PCO2 and PO2: a multicenter study of accuracy.

作者信息

Palmisano B W, Severinghaus J W

机构信息

Department of Anesthesia, Medical College of Wisconsin, Milwaukee.

出版信息

J Clin Monit. 1990 Jul;6(3):189-95. doi: 10.1007/BF02832145.

Abstract

A multicenter study used 756 samples from 251 patients in 12 institutions to compare arterial (PaO2, PaCO2) with transcutaneous (PsO2, PsCO2) oxygen and carbon dioxide tensions, measured usually at 44 degrees C. Of these samples, 336 were obtained from 116 neonates, 27 from 25 children with cystic fibrosis, and 140 from 40 patients under general anesthesia. Ninety-one patients were between 4 weeks and 18 years of age, 32 were between 18 and 60 years, and 12 were over 60. The ratio of transcutaneous to arterial P(s/a)CO2 was 1.01 +/- 0.11 with PaCO2 less than 30 mm Hg, increasing to 1.04 +/- 0.08 at PaCO2 greater than 40 mm Hg. Mean bias and its standard deviation (PsCO2 - PaCO2) were + 1.3 +/- 3.9 mm Hg in the entire group, + 1.8 +/- 4.2 mm Hg in neonates (NS). Bias was + 0.2 +/- 2.7 mm Hg when PaCO2 was less than 30 mm Hg (N = 175, NS), 1.0 +/- 3.4 with 30 less than PaCO2 less than 40 (n = 329, p less than 0.001), and + 2.04 +/- 4.00 mm Hg with 40 less than PaCO2 less than 70 (n = 229, p less than 0.001). These data suggest that, using transcutaneous PCO2 monitors with inbuilt temperature correction of 4.5%/degrees C, the skin metabolic offset should be set to 6 mm Hg. The linear regression was PsCO2 = 1.052(PaCO2) - 0.56, Sy.x = 3.92, R = 0.929 (n = 756); and PsCO2 = 1.09(PaCO2) - 1.57, Sy.x = 4.17, R = 0.928 in neonates (n = 336).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验