Orbscan II 和 Galilei G4 测量 Kappa 角:调节的影响。

Measurement of angle Kappa with Orbscan II and Galilei G4: effect of accommodation.

机构信息

Optometry Research Group (GIO), Department of Optics, University of Valencia, C/Dr Moliner, 50, 46100, Burjassot, Spain,

出版信息

Graefes Arch Clin Exp Ophthalmol. 2014 Feb;252(2):249-55. doi: 10.1007/s00417-013-2509-y. Epub 2013 Nov 20.

Abstract

PURPOSE

To measure angle kappa distance with Orbscan II and Galilei G4, and to evaluate possible variations in this value for different accommodation degrees.

METHODS

Angle kappa was measured using the Orbscan II and the Galilei G4 in the right eye of 80 patients aged from 20 to 40 years. This value was measured three times per eye and per device, and the average was retained. Angle kappa was measured for far vision using the Orbscan II and the Galilei G4 in a random order for each subject. The Galilei G4 was used to measure angle kappa as a function of accommodation, where the first measure started at +1 D and the vergence was changed until reaching -4 D, in 1 D steps. In both measures, the kappa distance was expressed in millimetres.

RESULTS

At distance, the values of angle kappa were 0.43 ± 0.13 mm and 0.27 ± 0.15 mm measured with the Orbscan II and Galilei G4 systems respectively. Statistical significant differences were found (P < 0.01). With regard to the angle kappa values obtained as a function of accommodation, the values were 0.25 ± 0.15 mm, 0.26 ± 0.15 mm, 0.30 ± 0.20 mm, 0.27 ± 0.15 and 0.26 ± 0.15 mm, for +1 D, -1 D, -2 D, -3 D and -4 D respectively. No statistical significant differences were found among 0 D and the other vergences evaluated (P > 0.01).

CONCLUSIONS

For far vision, Orbscan II measured significantly higher angle kappa values than Galilei G4, the mean difference being 0.16 ± 0.08 mm. For different accommodation levels, the kappa distance did not change significantly.

摘要

目的

使用 Orbscan II 和 Galilei G4 测量角kappa 距离,并评估在不同调节程度下该值可能发生的变化。

方法

使用 Orbscan II 和 Galilei G4 对 80 名年龄在 20 至 40 岁的患者的右眼进行角kappa 测量。每只眼和每种设备测量 3 次,保留平均值。使用 Orbscan II 和 Galilei G4 以随机顺序为每位受试者测量远视力时的角kappa。Galilei G4 用于测量作为调节函数的角kappa,其中第一次测量从+1 D 开始,并且直到达到-4 D 时改变收敛,以 1 D 为步长。在这两种测量中,kappa 距离均以毫米表示。

结果

在远距时,使用 Orbscan II 和 Galilei G4 系统分别测量的角kappa 值为 0.43 ± 0.13 mm 和 0.27 ± 0.15 mm。发现统计学上有显著差异(P < 0.01)。关于作为调节功能获得的角kappa 值,值分别为 0.25 ± 0.15 mm、0.26 ± 0.15 mm、0.30 ± 0.20 mm、0.27 ± 0.15 mm 和 0.26 ± 0.15 mm,+1 D、-1 D、-2 D、-3 D 和-4 D。0 D 和评估的其他收敛之间没有发现统计学上的显著差异(P > 0.01)。

结论

对于远距视力,Orbscan II 测量的角kappa 值明显高于 Galilei G4,平均差异为 0.16 ± 0.08 mm。对于不同的调节水平,kappa 距离没有明显变化。