使用Spectralis频域光学相干断层扫描技术对健康受试者进行视网膜神经纤维层测量时,手动和自动定中心的可重复性研究。

Reproducibility of retinal nerve fiber layer measurements with manual and automated centration in healthy subjects using spectralis spectral-domain optical coherence tomography.

作者信息

Lange Alex P, Sadjadi Reza, Costello Fiona, Guber Ivo, Traboulsee Anthony L

机构信息

Department of Ophthalmology and MS Clinic, Department of Neurology, University of British Columbia, Vancouver, BC, Canada V5Z 3N9.

MS Clinic, Department of Neurology, The University of British Columbia, Vancouer, BC, Canada V6T 2B5.

出版信息

ISRN Ophthalmol. 2012 Jul 30;2012:860819. doi: 10.5402/2012/860819. eCollection 2012.

Abstract

Objective. The aim of this study was to test the reproducibility of the Heidelberg Spectralis SD-OCT and to determine if provided software retest function for follow-up exam is superior to manual centration. Design. Prospective, cross-sectional study. Participants. 20 healthy subjects. Methods. All subjects underwent SD-OCT testing to determine retinal nerve fiber layer (RNFL) measurements sequentially on two different days and with two different centration techniques. Within-subject standard deviation, coefficient of variation, and intraclass correlation coefficient were used to assess reproducibility. Results. RNFL measurements showed high reproducibility, low within-subject standard deviation (1.3), low coefficient of variation (0.63%), and low intra-class correlation coefficient (0.98 (95% CI 0.97-0.99)) in the automated centration and manual centration groups for average RNFL Thickness. Quadrants showed slightly higher variability in the manual group compared to the automated group (within-subject standard deviation 2.5-5.3 versus 1.1-2.4, resp.). Conclusions. SD-OCT provides high-resolution RNFL measurements with high reproducibility and low variability. The re-test function allows for easier recentration for longitudinal examinations with similar results in average RNFL, but less variability in quadrant RNFL. SD-OCT high reproducibility and low variability is a promising fact and should be further evaluated in longitudinal studies of RNFL.

摘要

目的。本研究旨在测试海德堡Spectralis SD-OCT的可重复性,并确定其提供的用于随访检查的软件复测功能是否优于手动定中心。设计。前瞻性横断面研究。参与者。20名健康受试者。方法。所有受试者在两天内采用两种不同的定中心技术依次接受SD-OCT测试,以确定视网膜神经纤维层(RNFL)测量值。采用受试者内标准差、变异系数和组内相关系数来评估可重复性。结果。在自动定中心和手动定中心组中,平均RNFL厚度的RNFL测量显示出高可重复性、低受试者内标准差(1.3)、低变异系数(0.63%)和低组内相关系数(0.98(95%CI 0.97 - 0.99))。与自动定中心组相比,手动定中心组各象限的变异性略高(受试者内标准差分别为2.5 - 5.3和1.1 - 2.4)。结论。SD-OCT提供具有高可重复性和低变异性的高分辨率RNFL测量值。复测功能便于在纵向检查中重新定中心,平均RNFL的结果相似,但象限RNFL的变异性较小。SD-OCT的高可重复性和低变异性是一个有前景的事实,应在RNFL的纵向研究中进一步评估。