Several methods of measuring central corneal thickness (CCT) have been compared, however, the data are still limited and conflicting. The aim of this study was to determine the agreement of CCT measurements performed in healthy eyes using ultrasound pachymetry (USP), non-contact tono/pachymetry, specular microscopy, biometry, Scheimpflug-based corneal topography, and optical coherence tomography (OCT).
All of the participants underwent a complete ophthalmological examination. The CCT of all of the eyes included was measured using 6 different methods. The agreement between the methods was analyzed using the mean difference and Bland-Altman analysis based on a 95% limits of agreement.
A total of 64 patients with a mean age of 40.96±14.52 years (range: 20-78 years) were included in the study. The mean CCT value was 552.10±35.65 µm, 550.40±35.55 µm, 554.67±35.49 µm, 545.39±34.21 µm, 546.25±35.49 µm, and 552.64±33.59 µm using USP, non-contact tono/pachymetry, non-contact specular microscopy, biometry, Scheimpflug-based corneal topography, and OCT, respectively. The bias values determined by Bland-Altman plots were -1.70, -2.56, 6.71, 5.85, and -0.53 for tono/pachymetry, specular microscopy, biometry, topography, and OCT, respectively. OCT demonstrated the lowest bias compared to USP. The overall intraclass correlation coefficient was 0.961 (range: 0.945-0.974) with a 95% confidence interval.
All of the CCT measurements obtained using non-contact tono/pachymetry, non-contact specular microscopy, biometry, Scheimpflug-based corneal topography, and OCT were consistent with the USP measurements of healthy controls. Larger prospective studies to determine the interchangeability of different methods for CCT measurements in pathological conditions are warranted.
已对几种测量中央角膜厚度(CCT)的方法进行了比较,然而,数据仍然有限且相互矛盾。本研究的目的是确定使用超声测厚法(USP)、非接触眼压/测厚法、镜面显微镜检查、生物测量法、基于Scheimpflug原理的角膜地形图以及光学相干断层扫描(OCT)对健康眼睛进行CCT测量的一致性。
所有参与者均接受了全面的眼科检查。使用6种不同方法测量了所有纳入眼睛的CCT。基于95%一致性界限,采用平均差异和Bland-Altman分析对各方法之间的一致性进行了分析。
本研究共纳入64例患者,平均年龄为40.96±14.52岁(范围:20 - 78岁)。使用USP、非接触眼压/测厚法、非接触镜面显微镜检查、生物测量法、基于Scheimpflug原理的角膜地形图以及OCT测量得到的平均CCT值分别为552.10±35.65 µm、550.40±35.55 µm、554.67±35.49 µm、545.39±34.21 µm、546.25±35.49 µm和552.64±33.59 µm。Bland-Altman图确定的眼压/测厚法、镜面显微镜检查、生物测量法、角膜地形图以及OCT的偏差值分别为-1.70, -2.56, 6.71, 5.85和-0.53。与USP相比,OCT显示出最低的偏差。总体组内相关系数为0.961(范围:0.945 - 0.9