三种不同附着设计对上颌中切牙热塑性矫治器产生的垂直向力的影响。

Effect of three different attachment designs in the extrusive forces generated by thermoplastic aligners in the maxillary central incisor.

作者信息

Costa Rafael, Calheiros Fernanda Calabró, Ballester Rafael Yagüe, Gonçalves Flávia

机构信息

Faculdade de Odontologia, Universidade Ibirapuera, São Paulo, SP, Brazil.

Faculdade de Odontologia, Universidade de São Paulo, São Paulo, SP, Brazil.

出版信息

Dental Press J Orthod. 2020 May;25(3):46-53. doi: 10.1590/2177-6709.25.3.046-053.oar. Epub 2020 Aug 19.

Abstract

INTRODUCTION

Orthodontic aligners use have increased in dentistry. The resolution of complex movements such as extrusion demands the use of attachments to reach the aimed force, but just a few studies have been developed to evaluate the biomechanical performance of the aligners and their accessories.

OBJECTIVE

The objective of this study was to evaluate on the three axes (X, Y and Z) the forces generated by three different attachment designs for the extrusion of the maxillary central incisor using esthetic orthodontic aligners.

METHODS

Three prototypes of maxillary models were developed, each one with a specific attachment inserted in the central incisor. Three aligners were manufactured for each of the three attachment designs, with 0.33-mm activation in the direction of the extrusion. An analytical device was used to evaluate the forces applied to the three axes by each aligner/attachment. The data were assessed by one-way ANOVA and Tukey's test (α = 0.05).

RESULTS

All of the studied attachment designs could satisfactorily perform the extrusion movement. However, force intensities were different in the three designs (design 1 = 2.5 N; design 2 = 2.2 N, and design 3 = 1.1 N). Furthermore, two of the three attachment designs (designs 1 and 2) eventually exerted significant forces on the X (mesiodistal) and Y (buccopalatal) axes.

CONCLUSION

The attachment design 3 presents the best distribution of forces for extrusion movement, generating almost null forces on X and Y axes, and lower intensity of force on the Z axis.

摘要

引言

正畸矫治器在牙科领域的使用有所增加。诸如牙齿伸长等复杂移动的矫治需要使用附件来达到目标力,但仅有少数研究对矫治器及其附件的生物力学性能进行评估。

目的

本研究的目的是使用美观正畸矫治器,在三个轴(X、Y和Z)上评估三种不同附件设计对上颌中切牙伸长所产生的力。

方法

制作了三个上颌模型原型,每个模型的中切牙处插入一种特定附件。针对三种附件设计中的每一种制作三个矫治器,在伸长方向上激活0.33毫米。使用一种分析装置评估每个矫治器/附件在三个轴上施加的力。数据通过单因素方差分析和Tukey检验进行评估(α = 0.05)。

结果

所有研究的附件设计都能令人满意地完成伸长移动。然而,三种设计中的力强度不同(设计1 = 2.5牛;设计2 = 2.2牛,设计3 = 1.1牛)。此外,三种附件设计中的两种(设计1和2)最终在X轴(近远中)和Y轴(颊腭向)上施加了显著的力。

结论

附件设计3在伸长移动中呈现出最佳的力分布,在X轴和Y轴上产生几乎为零的力,在Z轴上的力强度较低。