Tañá-Rivero Pedro, Tello-Elordi Cristina, Orts-Vila Paz, Tañá-Sanz Pedro, Tañá-Sanz Santiago
Cataract Department, Oftalvist Alicante, Alicante, Spain.
Clin Ophthalmol. 2023 Apr 29;17:1245-1253. doi: 10.2147/OPTH.S409132. eCollection 2023.
To assess the agreement in automated corneal diameter (CD) and anterior chamber depth (ACD, measured from corneal epithelium to lens) distances measurements between the IOLMaster 500 and 700 optical biometers in myopic eyes.
One hundred and sixteen eyes of 116 myopic patients aged between 21 to 60 years were included in this study. Measurements of CD and ACD distances were taken with each biometer for all patients in the same session. A -test and a -value less than 0.05 was considered statistically significant when the measurements were compared. The agreement between biometers was studied by applying a Bland-Altman analysis.
The mean CD values obtained using the IOLMaster 500 and 700 biometers were 12.26±0.35 mm and 12.13±0.34 mm, respectively. The mean ACD distance values were 3.61±0.29 mm and 3.62±0.31, for the IOLMaster 500 and 700 biometers, respectively. There were statistically significant differences between the two devices only for CD measurements (<0.001), but not for ACD measurements (=0.26). The limits of agreement obtained were wide in both types of measurements, being 0.422 mm for the CD distance and 0.389 mm for the ACD distance.
There were statistically significant differences between the IOLMaster 500 and 700 biometers regarding CD but not in ACD measurements in healthy myopic eyes. These differences could be clinically significant in some cases. According to these results, medical judgement should be used to assess whether the two devices could be used interchangeably for CD and ACD measurements during the clinical practice.
评估IOLMaster 500和700光学生物测量仪在测量近视眼自动角膜直径(CD)和前房深度(ACD,从角膜上皮至晶状体测量)距离时的一致性。
本研究纳入了116例年龄在21至60岁之间的近视患者的116只眼。在同一时间段内,使用每台生物测量仪对所有患者进行CD和ACD距离测量。比较测量值时,采用t检验,P值小于0.05被认为具有统计学意义。通过应用Bland-Altman分析研究生物测量仪之间的一致性。
使用IOLMaster 500和700生物测量仪获得的平均CD值分别为12.26±0.35mm和12.13±0.34mm。IOLMaster 500和700生物测量仪的平均ACD距离值分别为3.61±0.29mm和3.62±0.31mm。仅在CD测量方面,两种设备之间存在统计学显著差异(P<0.001),而在ACD测量方面无差异(P=0.26)。在两种测量类型中获得的一致性界限都较宽,CD距离为0.422mm,ACD距离为0.389mm。
在健康近视眼中,IOLMaster 500和700生物测量仪在CD测量方面存在统计学显著差异,但在ACD测量方面无差异。在某些情况下,这些差异可能具有临床意义。根据这些结果,在临床实践中,应运用医学判断来评估这两种设备在CD和ACD测量中是否可互换使用。