To comprehensively assess the reliability of a new optical biometer (IOLMaster 700), based on swept-source optical coherence tomography (SS-OCT) and comparison with a standard biometer (IOLMaster 500), in healthy children, adults, and cataract patients.
A total of 301 eyes from 301 consecutive subjects were enrolled prospectively. Two experienced operators measured each eye three times consecutively with the IOLMaster 700. The axial length (AL), keratometry (K), anterior chamber depth (ACD), lens thickness (LT), central corneal thickness (CCT), and white-to-white (WTW) distance were recorded. Intraoperator repeatability and interoperator reproducibility of the IOLMaster 700 were analyzed using the test-retest (TRT), coefficients of variation (CoV), and intraclass correlation coefficients (ICCs). The agreement between the two devices was evaluated using the Bland-Altman method.
The repeatability and reproducibility of the SS-OCT optical biometer were high for all ocular biometry parameters in all groups, except for the WTW in cataract patients (TRT, 0.27-0.44 mm; ICC, 0.86-0.95). The reproducibility of averaged measurements from three consecutive readings (TRT : AL = 0.02 mm, CCT = 5.41 m, ACD = 0.03 mm, LT = 0.03 mm, Km = 0.17 D, and WTW = 0.22 mm) was higher than the reproducibility of single measurements (TRT : AL = 0.04 mm, CCT = 7.43 m, ACD = 0.06 mm, LT = 0.05 mm, Km = 0.26 D, and WTW = 0.35 mm) in the three groups. The consistency in the data between the two biometers was high, with narrow 95% LoAs in the three groups.
Repeatability and reproducibility of the new SS-OCT optical biometer were excellent and consistent with that of the standard biometer with respect to healthy children, healthy adults, and cataract patients.
基于扫频光学相干断层扫描(SS-OCT)全面评估新型光学生物测量仪(IOLMaster 700)在健康儿童、成人及白内障患者中的可靠性,并与标准生物测量仪(IOLMaster 500)进行比较。
前瞻性纳入301例连续受试者的301只眼。两名经验丰富的操作人员使用IOLMaster 700对每只眼连续测量3次。记录眼轴长度(AL)、角膜曲率(K)、前房深度(ACD)、晶状体厚度(LT)、中央角膜厚度(CCT)和白对白(WTW)距离。使用重测法(TRT)、变异系数(CoV)和组内相关系数(ICC)分析IOLMaster 700的同一操作人员重复性和不同操作人员再现性。使用Bland-Altman方法评估两种设备之间的一致性。
除白内障患者的WTW外,SS-OCT光学生物测量仪在所有组的所有眼生物测量参数方面的重复性和再现性均较高(TRT,0.27 - 0.44 mm;ICC,0.86 - 0.95)。三组中连续三次读数的平均测量值的再现性(TRT:AL = 0.02 mm,CCT = 5.41 m,ACD = 0.03 mm,LT = 0.03 mm,Km = 0.17 D,WTW = 0.22 mm)高于单次测量的再现性(TRT:AL = 0.04 mm,CCT = 7.43 m,ACD = 0.06 mm,LT = 0.05 mm,Km = 0.26 D,WTW = 0.35 mm)。两种生物测量仪之间的数据一致性较高,三组的95%一致性界限较窄。
新型SS-OCT光学生物测量仪在健康儿童、健康成人及白内障患者中的重复性和再现性极佳,且与标准生物测量仪一致。