Accuracy of optical coherence tomography (OCT) in pachymetry for glaucoma patients.

作者信息

Ayala Marcelo, Strandås Robert

机构信息

Eye Department, Skaraborgs Hospital, Skövde, Sahlgrenska Academy, Gothenburg University & Karolinska Institute, 541 85, Skövde, Sweden.

Eye Department, Skaraborgs Hospital, Skövde, Sweden.

出版信息

BMC Ophthalmol. 2015 Sep 29;15:124. doi: 10.1186/s12886-015-0116-x.

Abstract

BACKGROUND

Central corneal thickness (CCT) measurement has become an important test in the diagnosis and management of glaucoma. Currently, ultrasound corneal thickness measurement (pachymetry) is the most frequently used clinical technique and the gold standard to assess CCT. Newer instruments are currently available including the optical coherence tomography (OCT) instrument. The aim of the present study was therefore to evaluate the accuracy of the CCT measurements performed by three different observers, both with the OCT and ultrasound pachymetry (USP), in patients suffering from glaucoma.

METHODS

Patients who had been previously diagnosed with glaucoma participated in this cross-sectional study. Glaucoma was defined as patients who had at least two repeatable Humphrey visual fields showing glaucoma damage using the software 24-2, and with the optic nerve showing typical glaucoma damage. The patients CCTs were measured with OCT and USP by three different examiners.

RESULTS

Seventy eyes of 35 patients were included. The average age was 74 ± standard deviation (SD) 10.88, the average pachymetry value with OCT was 536 ± 29 μm, and the average pachymetry with USP was 532 ± 32 μm. The differences between OCT and USP were not significant (t-test, p = 0.32). The intraclass correlation coefficients were, for OCT, 0.99 [confidence interval (CI): 0.98-0.996], and for USP, 0.97 (CI: 0.95-0.98).

CONCLUSIONS

Agreement among the three observers using OCT or USP for pachymetry measurements was good. OCT might be used as an alternative method for pachymetry in glaucoma patients.

摘要

背景

中央角膜厚度(CCT)测量已成为青光眼诊断和治疗中的一项重要检查。目前,超声角膜厚度测量(测厚法)是评估CCT最常用的临床技术和金标准。目前已有包括光学相干断层扫描(OCT)仪器在内的更新型仪器。因此,本研究的目的是评估三位不同观察者使用OCT和超声测厚法(USP)对青光眼患者进行CCT测量的准确性。

方法

先前被诊断为青光眼的患者参与了这项横断面研究。青光眼定义为使用24-2软件至少有两次可重复的Humphrey视野显示青光眼损害,且视神经显示典型青光眼损害的患者。由三位不同的检查者使用OCT和USP测量患者的CCT。

结果

纳入了35例患者的70只眼。平均年龄为74±标准差(SD)10.88,OCT测量的平均角膜厚度值为536±29μm,USP测量的平均角膜厚度为532±32μm。OCT和USP之间的差异不显著(t检验,p = 0.32)。OCT的组内相关系数为0.99 [置信区间(CI):0.98 - 0.996],USP的组内相关系数为0.97(CI:0.95 - 0.98)。

结论

三位观察者使用OCT或USP进行角膜厚度测量的一致性良好。OCT可作为青光眼患者角膜厚度测量的替代方法。