科美康K2仪器的有效性和可靠性。

Validity and reliability of the Cosmed K2 instrument.

作者信息

Lucía A, Fleck S J, Gotshall R W, Kearney J T

机构信息

Sports Science Division, U.S. Olympic Committee, Colorado Springs.

出版信息

Int J Sports Med. 1993 Oct;14(7):380-6. doi: 10.1055/s-2007-1021196.

Abstract

The purpose of this study was to measure the validity and reliability of a portable telemetry system (Cosmed K2) for measurement of oxygen consumption (VO2) during a laboratory exercise test at submaximum and maximum intensities. Twenty subjects performed three testing sessions on a treadmill on three consecutive days. Each testing session consisted of a submaximal and a maximal exercise test. On one day, VO2 was determined with the conventional Douglas bag method, and on the other two days, VO2 was determined using the Cosmed K2 system. Utilization of the Douglas bag method or the Cosmed K2 was randomly assigned. The results indicated that the Cosmed K2 instrument is reliable and valid (p > 0.05) for measurement of VO2 (l.min-1) during laboratory exercise testing at submaximum and maximum intensities. The assumption made by the K2 apparatus of a constant respiratory gas exchange ratio (R) of 1.00 did not have a significant influence on VO2 measurements.

摘要

本研究的目的是测量一种便携式遥测系统(Cosmed K2)在亚最大强度和最大强度的实验室运动测试期间测量耗氧量(VO2)的有效性和可靠性。20名受试者连续三天在跑步机上进行了三次测试。每次测试包括一次亚最大运动测试和一次最大运动测试。一天,使用传统的道格拉斯袋法测定VO2,另外两天,使用Cosmed K2系统测定VO2。道格拉斯袋法或Cosmed K2的使用是随机分配的。结果表明,Cosmed K2仪器在亚最大强度和最大强度的实验室运动测试期间测量VO2(l.min-1)是可靠且有效的(p>0.05)。K2仪器假设呼吸气体交换率(R)恒定为1.00,这对VO2测量没有显著影响。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验