Chylack L T, Wolfe J K, Friend J, Khu P M, Singer D M, McCarthy D, del Carmen J, Rosner B
Center for Clinical Cataract Research, Brigham and Women's Hospital, Boston, Massachusetts.
Optom Vis Sci. 1993 Nov;70(11):886-95. doi: 10.1097/00006324-199311000-00005.
Subjective and objective systems are used to quantify cataract at The Center for Clinical Cataract Research. We have described each system and its use, presented data on reproducibility and validity, and for objective systems, demonstrated the correlation to the subjective grade of the cataract as defined by the Lens Opacities Classification Systems, Versions II and III (LOCS II and III). The subjective systems are used to classify nuclear color, nuclear opalescence, cortical cataract, and posterior subcapsular cataract. Reported kappa scores for LOCS II range from 0.85 to 1.0. Intraclass correlation coefficients for LOCS III (r1) range from 0.67 to 0.94. The computerized objective system are: (1) fast spectral scanning colorimetry (FSSC) for assessment of nuclear color (r1 = 0.96 to 0.98); (2) nuclear mean density (NMD) for assessment of nuclear opalescence (r1 = 0.97); and (3) percent area opacity (anterior = a; posterior = p) (OPAC-a and OPAC-p) for assessment of cortical and posterior subcapsular cataract (r1 = 0.92 to 0.96).
在临床白内障研究中心,主观和客观系统用于量化白内障。我们已经描述了每个系统及其用途,给出了关于可重复性和有效性的数据,对于客观系统,还展示了其与晶状体混浊分类系统第二版和第三版(LOCS II和III)所定义的白内障主观分级之间的相关性。主观系统用于对核颜色、核透明度、皮质性白内障和后囊下白内障进行分类。报告的LOCS II的kappa值范围为0.85至1.0。LOCS III(r1)的组内相关系数范围为0.67至0.94。计算机化客观系统包括:(1)用于评估核颜色的快速光谱扫描比色法(FSSC)(r1 = 0.96至0.98);(2)用于评估核透明度的核平均密度(NMD)(r1 = 0.97);以及(3)用于评估皮质性和后囊下白内障的面积混浊百分比(前部 = a;后部 = p)(OPAC-a和OPAC-p)(r1 = 0.92至0.96)。